文献速递妇产影像技术与安全US

改善肥胖孕妇中孕期胎儿超声图像质量:超声传播速度的作用

Improving image quality of mid-trimester fetal sonography in obese women: role of ultrasound propagation velocity.
PubMed 文献 · Ultrasound Obstet Gynecol · 2018-01-01

这项横断面研究纳入32例肥胖孕妇的中孕期胎儿超声检查图像,每例采集四个推荐扫查平面的图像组,每组包含在1540 m/s、1480 m/s和1420 m/s三种传播速度设置下连续获取的三幅图像;114名专家对100组图像按A至C分级评估质量。结果显示三种速度设置间的图像等级分布差异显著(P < 0.0001),1480 m/s时图像最常被评为A级,而常规1540 m/s时最常被评为C级;无论哪个扫查平面,腹壁脂肪层越厚,偏好的速度越低(P < 0.0001)。作者结论认为,采用低于1540 m/s的超声传播速度构建图像可显著改善肥胖孕妇中孕期胎儿超声图像质量。

要点

  • 肥胖孕妇中孕期胎儿超声采用1480 m/s传播速度时图像最常评为A级,常规1540 m/s最常评为C级(P < 0.0001)。
  • 腹壁脂肪层越厚,偏好的超声传播速度越低,各扫查平面均如此(P < 0.0001)。
  • 扫描仪默认按1540 m/s校准,而脂肪组织传播速度更慢(约1450 m/s),可影响图像质量。

核心结论

  • 采用低于1540 m/s的超声传播速度构建图像可显著改善肥胖孕妇中孕期胎儿超声图像质量,三种速度设置间图像等级分布差异显著(P < 0.0001)。

研究设计与样本

  • 横断面研究,纳入32例肥胖孕妇中孕期胎儿超声检查的100组图像,每组含1540 m/s、1480 m/s和1420 m/s三种传播速度下连续获取的图像,由114名专家按A至C分级评估。

诊断标准与关键数值

  • 图像质量按A(最可接受)至C(最不可接受)三级评定;1480 m/s时图像最常评为A级,1540 m/s时最常评为C级,三种速度设置间等级分布差异P < 0.0001。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The quality of ultrasound images is impaired in obese patients. All ultrasound scanners are calibrated for an ultrasound propagation velocity of 1540 m/s, but the propagation in fatty tissue is slower (in the order of 1450 m/s). The main objective of this study was to evaluate the quality of images obtained with different ultrasound propagation velocity settings during the mid-trimester fetal ultrasound examination in obese patients.

METHODS: This was a cross-sectional study using image sets of four recommended scanning planes collected from 32 obese pregnant women during their mid-trimester fetal scan. Each image set comprised three images obtained successively at three different propagation velocity settings (1540 m/s, 1480 m/s and 1420 m/s). A panel of 114 experts assessed the quality of 100 image sets, grading them from A (most acceptable) to C (least acceptable). Scanning-plane-specific indicators of adiposity (fatty layer thickness, probe-to-organ distance) were analyzed for each scanning plane.

RESULTS: The experts had a mean of 18.1 ± 10.2 years of experience. The grade distribution (A, B, C) differed significantly (P < 0.0001) between the three propagation velocity settings tested; at the lower speed of 1480 m/s, images were most often graded A, while at the conventional speed of 1540 m/s, they were most often graded C. Regardless of the scanning plane, the thicker the fatty layer of the abdominal wall in a given plane, the lower the preferred speed (P < 0.0001).

CONCLUSION: The construction of images taking into account ultrasound propagation velocities lower than 1540 m/s can improve significantly the quality of images obtained during mid-trimester fetal ultrasonography in obese women. Copyright © 2018 ISUOG. Published by John Wiley & Sons Ltd.

来源:Ultrasound Obstet Gynecol,查看原文。版权归原作者所有。