剪切波弹性成像评估肱动脉壁硬度:一种有前景的内皮功能障碍检测新诊断工具
前瞻性纳入65例参与者(19例经冠状动脉造影证实动脉粥样硬化、16例健康年轻成人、15例心血管危险因素患者、15例50至60岁健康老年人),同周行剪切波弹性成像与血流介导舒张检测。血流介导舒张值随危险因素和动脉粥样硬化而降低,弹性成像硬度变化值同步降低;作者认为剪切波弹性成像联合血流介导舒张可能是一种有前景、广泛可用的无创内皮功能障碍检测工具。
要点
- 剪切波弹性成像可测量肱动脉壁硬度变化,并与血流介导舒张同步变化。
- 健康年轻成人硬度变化19.9%±6.3%,动脉粥样硬化患者7.8%±6.4%,后者较健康老年人降低(P<.001)。
- 作者认为剪切波弹性成像联合血流介导舒张可能成为检测内皮功能障碍的无创工具。
核心结论
- 剪切波弹性成像联合血流介导舒张可能是一种有前景、广泛可用的无创内皮功能障碍检测工具。
研究设计与样本
- 前瞻性研究,连续纳入65例参与者,包括19例经冠状动脉造影证实动脉粥样硬化患者、16例健康年轻成人、15例心血管危险因素患者和15例50至60岁健康老年人,同周行剪切波弹性成像并评估血流介导舒张。
关键影像表现/征象
- 前臂阻断后血流介导舒张均值±2SD:健康年轻成人8.54%±1.4%,健康老年人7.61%±1.4%,危险因素患者5.83%±0.7%(P<.001),动脉粥样硬化患者3.81%±2.4%(相对危险因素组P<.001)。
- 硬度测量值随血流介导舒张增加而同步显著降低:健康年轻成人19.9%±6.3%,健康老年人16.3%±5.1%,危险因素患者9.8%±5.4%(相对无危险因素组P<.05),动脉粥样硬化患者7.8%±6.4%(相对健康老年人P<.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study was designed to measure the changes in brachial artery wall stiffness by shear wave elastography (SWE) and evaluate the accuracy of SWE changes for detection of endothelial dysfunction.
METHODS: Sixty-five consecutive participants (19 patients with atherosclerosis proven by coronary angiography, 16 healthy young adults, 15 patients with cardiovascular risk factors, and 15 healthy older adults between 50 and 60 years) were prospectively included in this study. They were examined in the same week by SWE, and flow-mediated dilatation was evaluated for each patient.
RESULTS: The mean flow-mediated dilatation values ± 2 SDs after forearm occlusion were 8.54% ± 1.4% in healthy young adults, 7.61% ± 1.4% in healthy older adults, 5.83% ± 0.7% in patients with risk factors (P < .001), and 3.81% ± 2.4% in patients with atherosclerosis (P < .001, with respect to the risk factor group). There was a significant decrease in stiffness measurements in parallel with the increase in flow-mediated dilatation: 19.9% ± 6.3% in healthy young adults, 16.3% ± 5.1% in healthy older adults, 9.8% ± 5.4% in patients with risk factors (P < .05 with respect to the group with no risk factors), and 7.8% ± 6.4% in patients with atherosclerosis (P < .001 with respect to the healthy older adults).
CONCLUSIONS: Shear wave elastography in combination with flow-mediated dilatation could be a promising, widely available noninvasive diagnostic tool for detecting endothelial dysfunction.