文献速递胸部技术与安全DR

点扩散函数反卷积算法提升图像质量的应用:胸部X线摄影的观察者偏好研究

Application of Deconvolution Algorithm of Point Spread Function in Improving Image Quality: An Observer Preference Study on Chest Radiography.
PubMed 文献 · Korean J Radiol · 2018-01-02

研究前瞻性纳入50对后前位胸部X线片,比较标准方案与附加TRUVIEW ART点扩散函数反卷积算法图像,由4名胸部放射科医师独立以5分偏好量表评价10个解剖区域显示度和整体图像质量,结果所有读者均偏好算法图像,所有区域及整体质量P<0.001。

要点

  • 点扩散函数反卷积算法可提升胸部X线片10个解剖区域及整体图像质量的观察者偏好。
  • 最受偏好的解剖区域为奇静脉食管隐窝、胸椎和未遮挡肺野。
  • 算法通过校正闪烁体产生的散射信号改善图像显示。

核心结论

  • 应用点扩散函数反卷积算法的胸部X线片解剖结构显示度优于原始胸部X线片。

研究设计与样本

  • 前瞻性研究,纳入50对后前位胸部X线片,分别采用标准方案和附加TRUVIEW ART算法采集,由4名胸部放射科医师独立评价。

诊断标准与关键数值

  • 采用5分偏好量表评价10个解剖区域显示度和整体图像质量;所有4名读者对全部10个解剖区域偏好算法图像(均值3.6,范围3.2-4.0,P<0.001),整体图像质量均值3.8,范围3.3-4.0,P<0.001。

临床应用与阅片要点

  • 阅片时可关注算法处理后的图像在奇静脉食管隐窝、胸椎和未遮挡肺野等区域的显示改善。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the preference of observers for image quality of chest radiography using the deconvolution algorithm of point spread function (PSF) (TRUVIEW ART algorithm, DRTECH Corp.) compared with that of original chest radiography for visualization of anatomic regions of the chest.

MATERIALS AND METHODS: Prospectively enrolled 50 pairs of posteroanterior chest radiographs collected with standard protocol and with additional TRUVIEW ART algorithm were compared by four chest radiologists. This algorithm corrects scattered signals generated by a scintillator. Readers independently evaluated the visibility of 10 anatomical regions and overall image quality with a 5-point scale of preference. The significance of the differences in reader's preference was tested with a Wilcoxon's signed rank test.

RESULTS: All four readers preferred the images applied with the algorithm to those without algorithm for all 10 anatomical regions (mean, 3.6; range, 3.2-4.0; p < 0.001) and for the overall image quality (mean, 3.8; range, 3.3-4.0; p < 0.001). The most preferred anatomical regions were the azygoesophageal recess, thoracic spine, and unobscured lung.

CONCLUSION: The visibility of chest anatomical structures applied with the deconvolution algorithm of PSF was superior to the original chest radiography.

来源:Korean J Radiol,查看原文。版权归原作者所有。