文献速递胸部技术与安全CT

矩阵大小对超高清CT肺部图像质量的影响:512×512、1024×1024与2048×2048的比较

Effect of Matrix Size on the Image Quality of Ultra-high-resolution CT of the Lung: Comparison of 512 × 512, 1024 × 1024, and 2048 × 2048.
PubMed 文献 · Acad Radiol · 2018-01-17

本研究在超高清CT上扫描狭缝体模和11个尸肺,比较512、1024、2048矩阵对空间分辨率和图像质量的影响。2048矩阵在320 mm视野下分辨率最高,对实性结节、磨玻璃影、肺气肿、小叶内网状影、蜂窝和血管清晰度的评分显著优于1024矩阵(P<.05);2048和1024矩阵除噪声和条纹伪影外均显著优于512矩阵(P<.001);视觉和定量噪声按512、1024、2048顺序显著降低(P<.001)。结论认为大矩阵在噪声增加的情况下仍保持空间分辨率并改善肺部疾病评估。

要点

  • 2048矩阵在320 mm视野下空间分辨率最高,对实性结节、磨玻璃影、肺气肿、小叶内网状影、蜂窝和血管清晰度的评分显著优于1024矩阵(P<.05)。
  • 2048和1024矩阵除噪声和条纹伪影外,整体图像质量和多数征象评分均显著优于512矩阵(P<.001)。
  • 视觉和定量噪声按512、1024、2048矩阵顺序显著降低(P<.001)。

核心结论

  • 在超高清CT中,与512矩阵相比,大矩阵可保持空间分辨率并改善图像质量和肺部疾病评估,尽管图像噪声增加。

研究设计与样本

  • 研究论文,使用狭缝体模和11个尸肺在超高清CT上扫描;体模采用20 mm视野1024矩阵及320 mm视野512、1024、2048矩阵重建,尸肺采用512、1024、2048矩阵重建;由3名观察者按3分法(1=最差,3=最佳)对整体图像质量、噪声、条纹伪影、血管、支气管和图像表现进行主观评分,采用Wilcoxon符号秩检验(Bonferroni校正)评估中位评分,噪声定量用Tukey检验,P<.05为显著。

诊断标准与关键数值

  • 最大空间分辨率为0.14 mm。
  • 320 mm视野下,2048矩阵分辨率最高,对整体质量、实性结节、磨玻璃影、肺气肿、小叶内网状影、蜂窝和血管清晰度的评分显著优于1024矩阵(P<.05)。
  • 2048和1024矩阵除噪声和条纹伪影外,均显著优于512矩阵(P<.001)。
  • 视觉和定量噪声按512、1024、2048矩阵顺序显著降低(P<.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: This study aimed to assess the effect of matrix size on the spatial resolution and image quality of ultra-high-resolution computed tomography (U-HRCT).

MATERIALS AND METHODS: Slit phantoms and 11 cadaveric lungs were scanned on U-HRCT. Slit phantom scans were reconstructed using a 20-mm field of view (FOV) with 1024 matrix size and a 320-mm FOV with 512, 1024, and 2048 matrix sizes. Cadaveric lung scans were reconstructed using 512, 1024, and 2048 matrix sizes. Three observers subjectively scored the images on a three-point scale (1 = worst, 3 = best), in terms of overall image quality, noise, streak artifact, vessel, bronchi, and image findings. The median score of the three observers was evaluated by Wilcoxon signed-rank test with Bonferroni correction. Noise was measured quantitatively and evaluated with the Tukey test. A P value of <.05 was considered significant.

RESULTS: The maximum spatial resolution was 0.14 mm; among the 320-mm FOV images, the 2048 matrix had the highest resolution and was significantly better than the 1024 matrix in terms of overall quality, solid nodule, ground-glass opacity, emphysema, intralobular reticulation, honeycombing, and clarity of vessels (P < .05). Both the 2048 and 1024 matrices performed significantly better than the 512 matrix (P < .001), except for noise and streak artifact. The visual and quantitative noise decreased significantly in the order of 512, 1024, and 2048 (P < .001).

CONCLUSION: In U-HRCT scans, a large matrix size maintained the spatial resolution and improved the image quality and assessment of lung diseases, despite an increase in image noise, when compared to a 512 matrix size.