比较基于剂量长度乘积与基于蒙特卡罗模拟器官计算的有效剂量在采用自动管电流调制的16层和64层MDCT检查中的差异

Comparing Dose-Length Product-Based and Monte Carlo Simulation Organ-Based Calculations of Effective Dose in 16- and 64-MDCT Examinations Using Automatic Tube Current Modulation.
PubMed 文献 · AJR Am J Roentgenol · 2018-01-30

本研究在16层和64层MDCT的头部、胸部及腹盆部CT和肺动脉CTA检查中,比较基于剂量长度乘积(DLP)计算的有效剂量(EDDLP)与蒙特卡罗模拟器官计算的有效剂量(EDMCO)。结果显示,EDDLP在男性及女性头部CT和女性胸部CT中显著低估EDMCO 19-39%,在男性腹盆部CT中高估12-14%,男性胸部CT和腹盆部CT偏差小于15%;16层与64层扫描仪间百分比差异无统计学显著差异。

要点

  • 基于DLP计算的有效剂量在头部CT和女性胸部CT中显著低估蒙特卡罗模拟器官剂量19-39%。
  • 男性腹盆部CT中DLP法高估有效剂量12-14%,男性胸部CT和腹盆部CT偏差小于15%。
  • 16层与64层MDCT扫描仪间有效剂量百分比差异无统计学显著差异(p≥0.13)。

核心结论

  • EDDLP在头部CT和女性胸部CT中低估EDMCO 19-39%,在男性胸部CT和腹盆部CT中偏差小于15%;差异归因于患者体型、自动管电流调制和性别中性k系数,计算ED时应考虑,尤其女性。

研究设计与样本

  • 研究纳入50例连续平扫头部CT、平扫胸部CT、平扫及增强腹盆部CT(使用16层和64层MDCT),50例64层MDCT肺动脉CTA和31例16层MDCT肺动脉CTA;用商用软件计算EDMCO和平均患者有效直径,并计算EDDLP;比较两种方法的平均差和百分比差异。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The purpose of this study is to compare dose-length product (DLP)-based calculation of effective dose (EDDLP) with Monte Carlo simulation organ-based calculation of effective dose (EDMCO) in 16- and 64-MDCT examinations, with the use of clinical examinations with automatic tube current modulation.

MATERIALS AND METHODS: Dose data were obtained from 50 consecutive unenhanced head CT examinations, unenhanced chest CT examinations, and unenhanced and contrast-enhanced abdominopelvic CT examinations performed using 16- and 64-MDCT scanners, as well as from 50 pulmonary CT angiography (CTA) examinations performed using a 64-MDCT scanner and 31 pulmonary CTA examinations performed using a 16-MDCT scanner. The EDMCO and the mean patient effective diameter were calculated using commercially available software. The EDDLP was also calculated. Both the mean difference and percentage difference between EDDLP and EDMCO were calculated, and they were statistically compared according to patient sex, type of examination performed, and type of scanner used.

RESULTS: EDDLP significantly underestimated the EDMCO by 0.3 mSv (19%) for men who underwent unenhanced head CT, 0.5 mSv (29%) for women who underwent unenhanced head CT, 0.9-1.4 mSv (9-13%) for men who underwent chest CT, and 4.7-4.8 mSv (39%) for women who underwent chest CT (p < 0.001). The EDDLP overestimated the EDMCO by 1.9-2.0 mSv (12-14%) for men who underwent abdominopelvic CT (p < 0.001), with no significant difference noted for women who underwent abdominopelvic CT's. No significant difference was noted in the percentage difference in ED between the 16- and 64-MDCT scanners (p ≥ 0.13).

CONCLUSION: EDDLP underestimates EDMCO, the reference standard for dose calculation, by 19-39% in unenhanced head CT examinations and, among women, in chest CT examinations. EDDLP deviates from EDMCO by less than 15% for chest CT examinations of men and for abdominopelvic CT. These differences can be attributed to variable patient body habitus, automatic tube current modulation, and sex-neutral k-coefficients, and they should be considered when calculating ED, particularly in women.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。