三级中心在Reveal LINQ植入式心脏监测系统存在下进行胸部CT和心脏MRI扫描的经验:关于伪影、患者安全和数据保存的病例系列回顾
在本三级心肺中心,对植入LINQ装置的患者行心脏MRI和胸部CT扫描,回顾性分析2014年3月至2016年12月间病例,由两名放射科医生评估皮下及胸腔内伪影及其临床意义,并审查LINQ数据下载和不良事件。所有扫描均见轻微皮下伪影,胸腔内伪影在胸部CT中占25.6%、心脏MRI中占33.3%,但无临床显著伪影;未见患者安全或数据丢失不良事件。
要点
- LINQ装置在所有胸部CT和心脏MRI扫描中均产生轻微皮下伪影,但无临床显著伪影。
- 胸腔内伪影见于25.6%的胸部CT和33.3%的心脏MRI,但均无临床意义。
- MRI扫描前后LINQ数据下载率分别为53.8%和56.5%,未观察到数据丢失事件。
核心结论
- LINQ装置不产生临床显著伪影,即使伪影延伸至胸腔内(见于约三分之一MRI和四分之一CT);MRI扫描LINQ装置是安全的,无证据显示不适当的数据丢失。
研究设计与样本
- 病例系列回顾;本地数据库检索2014年3月至2016年12月间接受胸部CT或心脏MRI的LINQ装置患者;由两名放射科医生审阅图像,记录皮下和胸腔内伪影及其临床意义;由心脏节律管理专家审阅MRI前后LINQ数据下载;检索信托事件报告系统。
关键影像表现/征象
- 所有扫描均存在轻微皮下伪影。
- 胸腔内伪影见于25.6%的胸部CT扫描和33.3%的心脏MRI扫描,但未观察到临床显著伪影。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: At our tertiary cardiothoracic centre, cardiac MRI and thoracic CT scans are performed in patients with implanted LINQ devices. The degree of foreign body artefact associated with the LINQ device, and its clinical importance, has not previously been assessed. A case series review was therefore performed with a simultaneous review of patient safety and data loss events, secondary to the MRI environment.
METHODS: A local database search identified LINQ device patients who underwent thoracic CT or cardiac MRI scans between March 2014 and December 2016. Images were reviewed by two radiologists, recording the presence of subcutaneous and intrathoracic artefact, and its clinical significance. Furthermore a specialist in cardiac rhythm management reviewed all LINQ data downloads undertaken before and after MRI scanning, and a search of the trust incident reporting system was performed.
RESULTS: Minor subcutaneous artefact was present on all scans. Intrathoracic artefact was observed in 25.6% of thoracic CT scans and 33.3% of cardiac MRIs; however no clinically significant artefact was observed. Device downloads were only performed by 53.8% of patients prior to their MRI scan and 56.5% after their MRI scan. No adverse patient safety or data loss events were noted.
CONCLUSION: The LINQ device does not produce clinically significant artefact, even when artefact extends into the intrathoracic space, which occurs in a third of MRIs and a quarter of CTs. MRI scanning of the LINQ device is safe with no evidence of inappropriate data loss. Advances in knowledge: This is the first published case series of CT and MRI scanning in LINQ patients and the first performed quantification of artefact related to the LINQ device.
引用
- Wiles BM, Illingworth CA, Couzins ML, et al. A tertiary centre experience of thoracic CT and cardiac MRI scanning in the presence of a reveal LINQ insertable cardiac monitoring system: a case series review of artefact, patient safety and data preservation. Br J Radiol. 2018. PMID: 29322834. doi:10.1259/bjr.20170615