双能CT单能外推联合迭代重建:减少颅内动脉瘤治疗中弹簧圈与夹子伪影
在体模模型中对14枚夹子和6枚弹簧圈行双能CT扫描,比较未校正、迭代金属伪影校正(MAR)、120 keV单能外推(VME)及MAR+VME联合四种重建。联合重建的伪影评分最低、邻近血管可见性改善最明显;对大型密集弹簧圈的严重伪影各技术均无改善。
要点
- 迭代MAR、VME及联合重建均较未校正图像显著降低夹子伪影并改善邻近血管可见性。
- 联合MAR+VME图像的伪影评分最低,优于单独使用任一技术。
- 对大型、密集填塞弹簧圈的严重伪影,所有技术均未显示改善。
核心结论
- 联合迭代MAR与VME可改善夹子及较小、疏松填塞弹簧圈的金属伪影校正;对大型且密集填塞的弹簧圈价值有限。
研究设计与样本
- 体模研究,在体模中扫描14枚夹子和6枚弹簧圈,采用双能CT在100与150SnkVp下采集,重建四组数据集:未校正滤波反投影、迭代MAR、双能CT 120 keV VME、迭代MAR+VME联合。两名独立盲法读片者行定性及定量评估。
关键影像表现/征象
- 与未校正图像相比,迭代MAR、VME及联合MAR+VME均显著降低定性夹子伪影(p < 0.001)和定量夹子伪影(p < 0.005),并改善邻近血管可见性(p < 0.05)。
- 联合迭代MAR+VME图像的伪影评分最低。
- 钛夹伪影较Phynox夹少(p < 0.05),且伪影评分随夹子尺寸增大而升高。
- 弹簧圈伪影随弹簧圈尺寸增大而增加,但迭代MAR+VME较未校正图像可减少伪影。
- 对大型、密集填塞弹簧圈的严重伪影,没有一种技术能改善。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare and to combine iterative metal artifact reduction (MAR) and virtual monoenergetic extrapolations (VMEs) from dual-energy computed tomography (DECT) for reducing metal artifacts from intracranial clips and coils.
METHODS: Fourteen clips and six coils were scanned in a phantom model with DECT at 100 and 150SnkVp. Four datasets were reconstructed: non-corrected images (filtered-back projection), iterative MAR, VME from DECT at 120 keV, and combined iterative MAR + VME images. Artifact severity scores and visibility of simulated, contrast-filled, adjacent vessels were assessed qualitatively and quantitatively by two independent, blinded readers.
RESULTS: Iterative MAR, VME, and combined iterative MAR + VME resulted in a significant reduction of qualitative (p < 0.001) and quantitative clip artifacts (p < 0.005) and improved the visibility of adjacent vessels (p < 0.05) compared to non-corrected images, with lowest artifact scores found in combined iterative MAR + VME images. Titanium clips demonstrated less artifacts than Phynox clips (p < 0.05), and artifact scores increased with clip size. Coil artifacts increased with coil size but were reducible when applying iterative MAR + VME compared to non-corrected images. However, no technique improved the severe artifacts from large, densely packed coils.
CONCLUSIONS: Combining iterative MAR with VME allows for an improved metal artifact reduction from clips and smaller, loosely packed coils. Limited value was found for large and densely packed coils.