跟腱剪切波弹性成像测量的可重复性
研究纳入14名健康志愿者,在足放松和固定于90°两种位置下测量跟腱剪切波速度,评估5次连续测量及连续5天的可重复性。结果示平均剪切波速度7.91–9.56 m/s,变异系数2.9%–6.3%,组内相关系数0.54–0.85;纵向扫描中放松位较固定位剪切波速度平均快0.47 m/s,放松位可重复性更高,操作者间组内相关系数横向0.70、纵向0.80。
要点
- 跟腱剪切波弹性成像在1小时及连续5天内均可获得可重复的剪切波速度测量。
- 纵向扫描采用放松足位时测量可重复性更高,放松位剪切波速度平均较固定位快0.47 m/s。
- 操作者间一致性较高,横向扫描组内相关系数0.70、纵向扫描0.80。
核心结论
- 跟腱剪切波弹性成像在1小时及连续5天内可获得可重复的测量,纵向平面联合放松足位测量更可靠,且操作者间一致性高,可作为定量评估人跟腱体内力学特性的可重复技术。
研究设计与样本
- 在体可重复性研究,纳入14名健康志愿者(男7名、女7名;平均年龄26.5±3.8岁,平均身高171.6±10.9 cm,平均VISA-A评分99.4±1.2),足放松及固定于90°条件下测量跟腱剪切波速度,采集连续5次测量及连续5天的数据。
诊断标准与关键数值
- 平均剪切波速度范围为7.91–9.56 m/s(±0.27–0.50 m/s)。
- 变异系数为2.9%–6.3%,相关系数为0.4–0.7,组内相关系数为0.54–0.85。
- 纵向扫描中不同足位间剪切波速度差异有统计学意义(p=<0.05),放松位平均较固定位快0.47 m/s。
- 操作者间组内相关系数横向扫描为0.70、纵向扫描为0.80。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To assess the reproducibility of shear wave elastography (SWE) measures in the Achilles tendon (AT) in vivo.
MATERIALS AND METHODS: Shear wave velocity (SWV) of 14 healthy volunteers [7 males, 7 females; mean age 26.5 ± 3.8 years, mean height 171.6 ± 10.9 cm, mean Victorian Institute of Sports Assessment Achilles questionnaire (VISA-A) score 99.4 ± 1.2] was measured with the foot relaxed and fixed at 90°. Data were collected over five consecutive measures and 5 consecutive days.
RESULTS: Mean SWV values ranged from 7.91 m/s-9.56 m/s ± 0.27-0.50 m/s. Coefficient of variation (CV), correlations and intra-class correlation coefficient (ICC) scores ranged from 2.9%-6.3%, 0.4-0.7 and 0.54-0.85 respectively. No significant differences were noted for longitudinal or transverse data with respect to protocol or time and no significant differences were noted for foot position in transverse data. Significant differences in SWV values were noted between foot positions for longitudinal scanning (p = <0.05), with a relaxed foot position providing SWV values on average 0.47 m/s faster than a fixed position. Increased reproducibility was obtained with the foot relaxed. ICC between operators was 0.70 for transverse and 0.80 for longitudinal scanning.
CONCLUSIONS: Reproducible SWE measures were obtained over a 1-h period as well as a period of 5 consecutive days with more reliable measures obtained from a longitudinal plane using a relaxed foot position. SWE also has a high level of agreement between operators making SWE a reproducible technique for quantitatively assessing the mechanical properties of the human AT in vivo.