乳腺扩散加权成像中的短T1反转恢复:与频谱衰减反转恢复在乳腺体模中的信噪比和表观扩散系数比较
本研究在3.0 T磁共振上使用含四种不同浓度砂糖琼脂凝胶的乳腺体模,比较短T1反转恢复(STIR-DWI)与频谱衰减反转恢复(SPAIR-DWI)两种脂肪抑制技术的信噪比和表观扩散系数。结果显示样本3和4的STIR-DWI信噪比低于SPAIR-DWI;样本2、3、4的STIR-DWI总体ADC显著高于SPAIR-DWI(均p<0.001),样本1无显著差异(p=0.62);STIR-DWI的ADC受T1值影响,T1值较短会降低信噪比、高估ADC并引入测量误差。
要点
- 样本3和4中STIR-DWI信噪比低于SPAIR-DWI。
- 样本2、3、4中STIR-DWI总体ADC显著高于SPAIR-DWI(均p<0.001)。
- STIR-DWI的ADC受T1值影响,T1值较短时信噪比降低、ADC被高估。
核心结论
- STIR-DWI获得的信噪比和ADC受T1值影响;T1值较短会降低信噪比、高估ADC并引入ADC测量误差。STIR-DWI在左右乳腺体模间的ADC差异大于SPAIR-DWI。
研究设计与样本
- 体模研究,乳腺体模由含四种不同浓度砂糖的琼脂凝胶组成(样本1、2、3、4),使用3.0 T磁共振行STIR-DWI和SPAIR-DWI并比较所得信噪比和ADC,同时比较左右乳腺体模间的ADC。
诊断标准与关键数值
- 样本3和4中STIR-DWI的信噪比低于SPAIR-DWI;样本2、3、4中STIR-DWI总体ADC显著高于SPAIR-DWI(均p<0.001),样本1无显著差异(p=0.62);Bland-Altman图显示STIR-DWI呈正向偏倚且一致性界限宽;STIR-DWI和SPAIR-DWI总体ADC的变异系数均良好;所有样本中STIR-DWI左右体模间ADC百分比差异略大于SPAIR-DWI。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: This study aimed to compare the signal-to-noise ratios (SNRs) and apparent diffusion coefficients (ADCs) obtained using two fat suppression techniques in breast diffusion-weighted imaging (DWI) of a phantom.
MATERIALS AND METHODS: The breast phantom comprised agar gels with four different concentrations of granulated sugar (samples 1, 2, 3, and 4). DWI with short tau inversion recovery (STIR-DWI) and that with spectral attenuated inversion recovery (SPAIR-DWI) were performed using 3.0-T magnetic resonance imaging, and the obtained SNRs and ADCs were compared. ADCs were also compared between the right and left breast phantoms.
RESULTS: For samples 3 and 4, SNRs obtained using STIR-DWI were lower than those obtained using SPAIR-DWI. For samples 2, 3, and 4, overall ADCs obtained using STIR-DWI were significantly higher than those obtained using SPAIR-DWI (p < 0.001 for all), although no significant difference was observed for sample 1 (p = 0.62). STIR-DWI shows a positive bias and wide limits of agreement in Bland-Altman plot. The coefficients of variance of overall ADCs were good in STIR-DWI and SPAIR-DWI. For all samples, STIR-DWI demonstrated slightly larger percentage differences in ADCs between the right and left phantoms than SPAIR-DWI.
CONCLUSION: SNRs and ADCs obtained using STIR-DWI are influenced by the T 1 value; a shorter T 1 value decreases SNRs, overestimates ADCs, and induces the measurement error in ADCs. STIR-DWI showed a larger difference in ADCs between the right and left phantoms than SPAIR-DWI.
引用
- Yoshida T, Urikura A, Shirata K, et al. Short tau inversion recovery in breast diffusion-weighted imaging: signal-to-noise ratio and apparent diffusion coefficients using a breast phantom in comparison with spectral attenuated inversion recovery. Radiol Med. 2017. PMID: 29230679. doi:10.1007/s11547-017-0840-9