不同工作站测量表观扩散系数可重复性的评估
这项前瞻性研究纳入100个经活检证实的乳腺病灶,在3 T MRI上采用b值0和850 s/mm²的扩散加权成像序列,由两名放射科医生共识勾画感兴趣区,比较随机工作站与商用DICOM查看器两种平台测量的平均ADC值。结果两种工作站测量一致性高(组内相关系数=0.981),Bland-Altman图未发现系统差异,结论为ADC测量在所考虑的工作站间可重复。
要点
- 乳腺病灶ADC测量在随机工作站与商用DICOM查看器间一致性高(组内相关系数=0.981)。
- Bland-Altman分析未发现两种工作站间存在系统差异。
- 一致性界限范围为最小-0.071×10⁻³ mm²/s至最大0.102×10⁻³ mm²/s。
核心结论
- ADC测量在本研究考虑的工作站间可重复。
研究设计与样本
- 前瞻性研究纳入100个经活检证实的乳腺病灶,MRI于3 T使用标准序列及b值0和850 s/mm²的回波平面扩散加权成像序列,由两名放射科医生共识阅片并沿病灶轮廓手动放置感兴趣区,排除邻近正常组织、坏死组织及囊性成分,分别在随机工作站和商用DICOM查看器上测量每个病灶的平均ADC值。
诊断标准与关键数值
- 两种工作站测量一致性高,组内相关系数=0.981;Bland-Altman图未发现系统差异,一致性界限范围为最小-0.071×10⁻³ mm²/s至最大0.102×10⁻³ mm²/s。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate apparent diffusion coefficient (ADC) measurements of breast lesions on different computer platforms to address post-processing influences on ADC measurement reproducibility.
MATERIALS AND METHODS: One hundred biopsy-proven breast lesions were included in this prospective study. MRI examination was performed at 3 T using standard sequences and an echo planar diffusion-weighted imaging sequence with b-values of 0 and 850 s/mm2. The images were reviewed by two radiologists in consensus. Regions of interest were placed manually within the lesion, following its contour. Care was taken to exclude adjacent normal tissue or necrotic tissue and cystic components within the lesion. The mean ADC value was measured for each lesion on two different platforms: On the MRI workstation that came with the scanner and on a commercially available DICOM (digital imaging and communication in medicine) viewer. Agreement between workstation measurements was evaluated using intraclass correlation coefficient and Bland-Altman plots.
RESULTS: Fifty-nine malignant and 41 benign lesions were analysed. Of the benign lesions, 28 were mass lesions and 13 were non-mass-like enhancements. In addition, 46 of the malignant lesions were masses and 13 were non-mass-like enhancements. Agreement between the two workstation measurements was high (intraclass correlation coefficients=0.981). Using Bland-Altman plots, no systematic differences were identified between workstations. Limits of agreement ranged between a minimum of -0.071×10-3 mm2/s and a maximum of 0.102×10-3 mm2/s.
CONCLUSION: ADC measurements are reproducible among the workstations considered in this study.