影响对比增强能谱乳腺摄影病灶尺寸测量精度的因素
回顾性分析204例乳腺癌,评估对比增强能谱乳腺摄影(CESM)术前测量肿瘤大小的精度,以病理显微最大径为参考标准。CESM平均高估2.9 mm(95%CI -10.3~16.2 mm),Spearman相关系数0.83(p<0.0001),37.8%测量一致、47%高估、15.2%低估;肿瘤大小、淋巴结受累、乳腺密度和乳腺大小显著影响测量准确性。
要点
- CESM测量乳腺癌大小平均高估2.9 mm,与病理相关性好(r=0.83)。
- 仅37.8%的CESM测量与病理一致,47%高估、15.2%低估。
- 肿瘤大小、淋巴结受累、乳腺密度和乳腺大小显著影响测量精度。
核心结论
- CESM预测肿瘤大小的质量良好,在乳腺癌手术规划中似乎是一种有前景的影像技术;但肿瘤生物学特征和患者人体测量学特征会影响诊断精度,应予以考虑。
研究设计与样本
- 回顾性研究,纳入204例乳腺癌,记录肿瘤生物学和人体测量学变量,以病理显微测量肿瘤最大径为参考标准。
关键影像表现/征象
- CESM测量肿瘤平均大小23.6±16.7 mm,病理平均大小20.7±15.8 mm。
- Bland-Altman分析显示CESM平均高估2.9 mm(95%CI -10.3~16.2 mm)。
- Spearman相关系数0.83(p<0.0001)。
- 测量一致性分析:37.8%一致,47%高估,15.2%低估。
诊断标准与关键数值
- 以病理显微测量肿瘤最大径为参考标准,CESM与病理尺寸的一致性定义为测量值相等。
- 肿瘤大小、淋巴结受累、乳腺密度和乳腺大小显著改变尺寸测量准确性。
临床应用与阅片要点
- CESM可用于乳腺癌术前肿瘤大小预测,但需注意其平均高估约2.9 mm。
- 报告时应结合肿瘤大小、淋巴结受累、乳腺密度和乳腺大小等因素综合评估测量精度。
- 作者认为肿瘤生物学特征和患者人体测量学特征影响诊断精度,应予以考虑。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the precision of the pre-surgical measurement of the size of breast cancer by contrast-enhanced spectral mammography (CESM).
MATERIAL AND METHODS: This was a retrospective study of 204 breast cancers. Variables related to tumour biology and anthropometric variables were recorded and considered when evaluating the efficacy of CESM in predicting tumour size. Microscopic measurement of the largest diameter of the tumour at pathology was chosen as the reference standard.
RESULTS: The mean size of tumours at pathology was 20.7±15.8 mm, while at CESM it was 23.6±16.7 mm (Bland-Altman 2.9 mm overestimation, 2.9 mm; 95% confidence interval [CI]: -10.3-16.2 mm). Spearman's correlation coefficient was 0.83 (p<0.0001). The concordance analysis indicated that 37.8% of the measurements were concordant, 47% were overestimated, and 15.2% were underestimated. Tumour size, nodal involvement, breast density, and breast size significantly modified the sizing accuracy.
CONCLUSION: Quality of tumour size prediction with CESM is good, and this appears to be a promising imaging technique in the surgical planning of breast cancer. Biological tumour features, and anthropological characteristics of the patients do, however, affect the diagnostic precision and should be taken into account.
引用
- Travieso-Aja MDM, Naranjo-Santana P, Fernández-Ruiz C, et al. Factors affecting the precision of lesion sizing with contrast-enhanced spectral mammography. Clin Radiol. 2017. PMID: 29221721. doi:10.1016/j.crad.2017.10.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。