阑尾CT能否低于1 mSv?一项降阶梯剂量模拟研究

Can We Perform CT of the Appendix with Less Than 1 mSv? A De-escalating Dose-simulation Study.
PubMed 文献 · Eur Radiol · 2017-12-07

前瞻性纳入30例疑似阑尾炎年轻患者,用低剂量模拟工具生成1.5、1.0、0.5 mSv CT,以2.0 mSv为参照做非劣效检验。腹部放射科医生对1.5、1.0、0.5 mSv均非劣效;非腹部放射科医生仅1.5、1.0 mSv非劣效,0.5 mSv未能证明非劣效。

要点

  • 腹部与非腹部放射科医生阅片时,1.0 mSv阑尾CT对2.0 mSv均达到非劣效。
  • 0.5 mSv仅对腹部放射科医生非劣效,对非腹部放射科医生未能证明非劣效。
  • 读者经验是低剂量CT诊断效能受损的重要影响因素。

核心结论

  • 1.0 mSv阑尾CT对腹部和非腹部放射科医生的诊断效能均非劣于2.0 mSv;0.5 mSv仅对腹部放射科医生非劣效。

研究设计与样本

  • 前瞻性纳入30例因疑似阑尾炎行2.0 mSv CT的年轻成人,用低剂量模拟工具和知识型迭代重建从螺旋投影数据生成1.5、1.0、0.5 mSv CT,非劣效界值0.06,主要终点为3名腹部和3名非腹部放射科医生的合并AUC。

诊断标准与关键数值

  • 腹部放射科医生2.0 vs 0.5 mSv合并AUC差0.017(95%CI -0.016, 0.050),1.5、1.0、0.5 mSv非劣效均被接受;非腹部放射科医生2.0 vs 1.0 mSv合并AUC差-0.017(95%CI -0.070, 0.035)非劣效被接受,2.0 vs 0.5 mSv合并AUC差0.045(95%CI -0.071, 0.161)非劣效未能证明。

临床应用与阅片要点

  • 对于腹部和非腹部放射科医生,1.0 mSv阑尾CT可能可行;0.5 mSv CT仅对经验丰富的腹部放射科医生非劣效于2.0 mSv;读者经验是低剂量CT诊断受损的重要因素。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To systematically explore the lowest reasonably achievable radiation dose for appendiceal CT using an iterative reconstruction (IR) in young adults.

METHODS: We prospectively included 30 patients who underwent 2.0-mSv CT for suspected appendicitis. From the helical projection data, 1.5-, 1.0- and 0.5-mSv CTs were generated using a low-dose simulation tool and the knowledge-based IR. We performed step-wise non-inferiority tests sequentially comparing 2.0-mSv CT with each of 1.5-, 1.0- and 0.5-mSv CT, with a predetermined non-inferiority margin of 0.06. The primary end point was the pooled area under the receiver-operating-characteristic curve (AUC) for three abdominal and three non-abdominal radiologists.

RESULTS: For the abdominal radiologists, the non-inferiorities of 1.5-, 1.0- and 0.5-mSv CT to 2.0-mSv CT were sequentially accepted [pooled AUC difference: 2.0 vs. 0.5 mSv, 0.017 (95% CI: -0.016, 0.050)]. For the non-abdominal radiologists, the non-inferiorities of 1.5- and 1.0-mSv CT were accepted; however, the non-inferiority of 0.5-mSv CT could not be proved [pooled AUC difference: 2.0 vs. 1.0 mSv, -0.017 (-0.070, 0.035) and 2.0 vs. 0.5 mSv, 0.045 (-0.071, 0.161)].

CONCLUSION: The 1.0-mSv appendiceal CT was non-inferior to 2.0-mSv CT in terms of diagnostic performance for both abdominal and non-abdominal radiologists; 0.5-mSv appendiceal CT was non-inferior only for abdominal radiologists.

KEY POINTS: • For both abdominal and non-abdominal radiologists, 1.0-mSv appendiceal CT could be feasible. • The 0.5-mSv CT was non-inferior to 2.0-mSv CT only for expert abdominal radiologists. • Reader experience is an important factor affecting diagnostic impairment by low-dose CT.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Park JH, Jeon JJ, Lee SS, et al. Can We Perform CT of the Appendix with Less Than 1 mSv? A De-escalating Dose-simulation Study. Eur Radiol. 2017. PMID: 29218613. doi:10.1007/s00330-017-5159-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。