基于模型的迭代重建在低辐射剂量CT结肠成像中的应用:结直肠癌术前评估
本研究纳入30例拟手术的结直肠癌患者,分别行标准剂量仰卧位和低剂量(辐射剂量降低约85%)俯卧位CTC,标准剂量图像用滤波反投影重建,低剂量图像用滤波反投影和基于模型的迭代重建重建。结果显示基于模型的迭代重建可显著改善低辐射剂量CTC的图像质量,并提供与标准辐射剂量方案相当的术前信息。
要点
- 低剂量CTC采用基于模型的迭代重建后,图像噪声较标准剂量滤波反投影更低(11.2 vs 17.3 HU)。
- 基于模型的迭代重建的低剂量CTC术前信息准确性优于低剂量滤波反投影,与标准剂量滤波反投影相当。
- 低剂量方案辐射剂量较标准剂量降低约85%,基于模型的迭代重建可弥补图像质量下降。
核心结论
- 基于模型的迭代重建可显著改善低辐射剂量CTC的图像质量,并提供与标准辐射剂量方案相当的术前信息。
研究设计与样本
- 研究纳入30例拟手术治疗的结直肠癌患者,所有患者均行标准剂量仰卧位CTC和低剂量俯卧位CTC(辐射剂量降低约85%),标准剂量图像用滤波反投影重建,低剂量图像用滤波反投影和基于模型的迭代重建重建,比较三种方法的客观和主观图像质量,并将术前评估(部位、长度、肿瘤和淋巴结分期)与术后病理结果对照。
关键影像表现/征象
- 标准剂量滤波反投影、低剂量滤波反投影和低剂量基于模型的迭代重建的平均图像噪声分别为17.3±3.2、40.5±10.9和11.2±2.0 HU,三种方法所有两两比较均有显著差异(P<.01)。
- 图像噪声的平均视觉评分:标准剂量滤波反投影为3.9±0.3,低剂量滤波反投影为2.0±0.5,低剂量基于模型的迭代重建为3.7±0.3;标准剂量滤波反投影和低剂量基于模型的迭代重建的评分显著高于低剂量滤波反投影,且两者相当。
- 标准剂量滤波反投影和低剂量基于模型的迭代重建的术前信息准确性优于低剂量滤波反投影,标准剂量滤波反投影与低剂量基于模型的迭代重建之间相当。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To assess the effect of model-based iterative reconstruction (MBIR) on image quality and diagnostic performance of low-radiation-dose computed tomography colonography (CTC) in the preoperative assessment of colorectal cancer.
MATERIALS AND METHODS: This study included 30 patients with colorectal cancer referred for surgical treatment. All patients underwent CTC with a standard dose (SD) protocol in the supine position and a low-dose (LD; radiation dose reduction of approximately 85%) protocol in the prone position. The SD protocol images were post-processed using filtered back projection (FBP), whereas the LD protocol images were post-processed using FBP and MBIR. Objective and subjective image quality parameters were compared among the three different methods. Preoperative evaluations, including site, length, and tumor and node staging were performed, and the findings were compared to the postsurgical findings.
RESULTS: The mean image noise of SD-FBP, LD-FBP, and LD-MBIR images was 17.3 ± 3.2, 40.5 ± 10.9, and 11.2 ± 2.0 Hounsfield units, respectively. There were significant differences for all comparison combinations among the three methods (P < .01). For image noise, the mean visual scores were significantly higher for SD-FBP and LD-MBIR than for LD-FBP, and the scores for SD-FBP and LD-MBIR were equivalent (3.9 ± 0.3 [SD-FBP], 2.0 ± 0.5 [LD-FBP], and 3.7 ± 0.3 [LD-MBIR]). Preoperative information was more accurate under SD-FBP and LD-MBIR than under LD-FBP, and the information was comparable between SD-FBP and LD-MBIR.
CONCLUSION: MBIR can yield significantly improved image quality on low-radiation-dose CTC and provide preoperative information equivalent to that of standard-radiation-dose protocol.