利用光谱探测器CT虚拟单能图像评估动脉期高强化肝脏病灶:体模与患者经验

Assessment of arterially hyper-enhancing liver lesions using virtual monoenergetic images from spectral detector CT: phantom and patient experience.
PubMed 文献 · Abdom Radiol (NY) · 2018-01-01

在体模和患者中研究虚拟单能重建(VMI)对动脉期高强化肝脏病灶评估的价值,并比较常规图像的混合迭代与光谱重建。结果显示VMI 40 keV使信噪比提高约三倍,病灶对比噪声比和病灶-肝脏比亦提高约三倍,主观病灶勾画最佳。作者认为VMI可改善此类病灶的评估。

要点

  • VMI 40 keV较常规图像使病灶对比噪声比和病灶-肝脏比提高约三倍。
  • VMI 40 keV的主观病灶勾画、可感知噪声和整体图像质量均最佳。
  • 常规混合迭代与光谱重建之间信噪比无差异(p=1)。

核心结论

  • VMI可改善动脉期高强化肝脏病灶的评估,因其在整个keV能谱范围内提高病灶对比度同时保持低图像噪声;作者认为可考虑在动脉期高强化肝脏病灶发现后行VMI筛查。

研究设计与样本

  • 体模与患者研究,在光谱探测器CT(Philips Healthcare, Best, The Netherlands)上完成全部成像;体模为非拟人体模,内含模拟病灶插入物(碘水溶液);患者为动脉期对比增强检查图像;重建VMI(40-200 keV,10 keV递增)、CI-IR和CI-SR,采用不同强度图像去噪;共评估20例患者的51个肝脏病灶,由两名放射科医生对40、70、110 keV和CI图像按五点Likert量表评分。

关键影像表现/征象

  • 体模中VMI 40 keV的信噪比较CI-IR/SR提高三倍(5.8 ± 1.1 vs. 18.8 ± 2.2,p ≤ 0.001)。
  • CI-IR与CI-SR之间信噪比无差异(p = 1)。
  • 去噪可降低噪声40%。
  • 患者中VMI 40 keV的衰减最高,图像噪声与CI-IR相当,对比噪声比/病灶-肝脏比提高三倍(CI-IR 1.3 ± 0.8/4.4 ± 2.0,VMI 40 keV:3.8 ± 2.7/14.2 ± 7.5,p ≤ 0.001)。

诊断标准与关键数值

  • 体模信噪比:CI-IR/SR 5.8 ± 1.1 vs. VMI 40 keV 18.8 ± 2.2,p ≤ 0.001。
  • 患者对比噪声比/病灶-肝脏比:CI-IR 1.3 ± 0.8/4.4 ± 2.0 vs. VMI 40 keV 3.8 ± 2.7/14.2 ± 7.5,p ≤ 0.001。
  • 主观病灶勾画VMI 40 keV最佳(p ≤ 0.01),并提供最低可感知噪声和最佳整体图像质量。
  • 去噪降低噪声40%。

临床应用与阅片要点

  • 动脉期高强化肝脏病灶评估中,VMI 40 keV可提高病灶对比度并保持低噪声,有助于病灶勾画。
  • 常规CI-IR与CI-SR信噪比无差异,选择时可参考其他因素。
  • 作者建议在动脉期高强化肝脏病灶发现后考虑VMI筛查。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To investigate a benefit from virtual monoenergetic reconstructions (VMIs) for assessment of arterially hyper-enhancing liver lesions in phantom and patients and to compare hybrid-iterative and spectral image reconstructions of conventional images (CI-IR and CI-SR).

METHODS: All imaging was performed on a SDCT (Philips Healthcare, Best, The Netherlands). Images of a non-anthropomorphic phantom with a lesion-mimicking insert (containing iodine in water solution) and arterial-phase images from contrast-enhanced patient examinations were evaluated. VMIs (40-200 keV, 10 keV increment), CI-IR, and CI-SR were reconstructed using different strengths of image denoising. ROIs were placed in lesions, liver/matrix, muscle; signal-to-noise, contrast-to-noise, and lesion-to-liver ratios (SNR, CNR, and LLR) were calculated. Qualitatively, 40, 70, and 110 keV and CI images were assessed by two radiologists on five-point Likert scales regarding overall image quality, lesion assessment, and noise.

RESULTS: In phantoms, SNR was increased threefold by VMI40keV compared with CI-IR/SR (5.8 ± 1.1 vs. 18.8 ± 2.2, p ≤ 0.001), while no difference was found between CI-IR and CI-SR (p = 1). Denoising was capable of noise reduction by 40%. In total, 20 patients exhibiting 51 liver lesions were assessed. Attenuation was the highest in VMI40keV, while image noise was comparable to CI-IR resulting in a threefold increase of CNR/LLR (CI-IR 1.3 ± 0.8/4.4 ± 2.0, VMI40keV: 3.8 ± 2.7/14.2 ± 7.5, p ≤ 0.001). Subjective lesion delineation was the best in VMI40keV image (p ≤ 0.01), which also provided the lowest perceptible noise and the best overall image quality.

CONCLUSIONS: VMIs improve assessment of arterially hyper-enhancing liver lesions since they increase lesion contrast while maintaining low image noise throughout the entire keV spectrum. These data suggest that to consider VMI screening after arterially hyper-enhancing liver lesions.