新型探测器基光谱CT虚拟单能图像噪声特征研究

Noise characteristics of virtual monoenergetic images from a novel detector-based spectral CT scanner.
PubMed 文献 · Eur J Radiol · 2017-11-10

该研究在体模和患者中评估新型双层探测器光谱CT(SDCT)虚拟单能图像(VMI)的噪声特征。体模噪声在40keV最高(5.3±0.2 HU),70keV后恒定;患者各部位VMI噪声均低(骨除外),肝脏、胰腺、门静脉、主动脉、肌肉和脂肪各能级VMI噪声均低于常规图像(p<0.01);SNR和CNR在40keV最高,随能量升高逐渐下降,120keV后恒定。VMI≤70keV定性评分显著高于常规图像(p≤0.01),≥80keV则低于常规图像(p<0.001)。

要点

  • VMI在40keV噪声最高,70keV后噪声保持恒定低水平。
  • 肝脏等腹部实质脏器各能级VMI噪声均低于常规120kVp图像(p<0.01)。
  • VMI≤70keV主观评分优于常规图像,≥80keV则劣于常规图像。

核心结论

  • 新型SDCT获得的VMI在全能量谱范围内噪声低,SNR和CNR较常规120kVp图像有显著改善。

研究设计与样本

  • 体模研究加患者研究,使用Catphan® 600圆柱体模行SDCT扫描,并连续纳入100例行腹部CT的SDCT患者,测量VMI(40-200keV)及常规120kVp图像的噪声、SNR和CNR,由独立读者对血管强化和图像质量行5分制定性分析(1最差,5最佳)。

关键影像表现/征象

  • 体模噪声在40keV最高(5.3±0.2 HU),逐渐降至70keV(3.6±0.2 HU),之后至200keV保持恒定(3.5±0.2 HU)。
  • 患者队列中除骨外所有区域各能级VMI噪声均低(<25 HU),肝脏噪声在40keV最高(13.2±4.6 HU),降至70keV(12.0±4.4 HU)后至200keV保持恒定低水平(11.6±4.3 HU)。
  • 肝脏、胰腺、门静脉、主动脉、肌肉和脂肪各能级VMI噪声均低于常规图像(p<0.01)。
  • 所有器官SNR和CNR在40keV最高(分别为6.8-34.9和18.3-44.9),逐渐降至120keV(分别为3.4-6.5和9.5-13.0),之后至200keV保持恒定(分别为2.6-5.5和8.5-12.5)。

诊断标准与关键数值

  • 定性分析采用5分制(1最差,5最佳),VMI≤70keV评分显著高于常规图像(p≤0.01),VMI≥80keV评分低于常规图像(p<0.001)。

临床应用与阅片要点

  • 低keV(≤70keV)VMI可提供更高的SNR、CNR及主观图像质量评分,高keV(≥80keV)VMI主观评分低于常规图像。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To evaluate the noise characteristics of virtual monoenergetic images (VMI) obtained from a recently introduced dual-layer detector-based spectral CT (SDCT), both in a phantom and patients.

MATERIALS AND METHODS: A cylindrical Catphan® 600 phantom (The Phantom Library, Salem NY, USA) was scanned using the SDCT. Image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured in VMI from 40 to 200keV as well as conventional 120 kVp images. One hundred consecutive patients who had an abdominal CT on the SDCT were then recruited in the study. Noise, SNR and CNR were measured in the liver, pancreas, spleen, kidney, abdominal aorta, portal vein, muscle, bone, and fat, both in VMI (40-200 keV) and conventional 120kVp images. Qualitative image analysis was performed by an independent reader for vascular enhancement and image quality on a 5 point scale (1-worst, 5-best).

RESULTS: On phantom studies, noise was low at all energies of VMI. Noise was highest at 40keV (5.3±0.2 HU), gradually decreased up to 70keV (3.6±0.2 HU), after which it remained constant up to 200keV (3.5±0.2 HU). In the patient cohort, noise was low (<25 HU) at all the energy levels of VMI for all the regions, with the exception of bone. For example, noise in the liver was highest at 40keV (13.2±4.6 HU), steadily decreased up to 70keV (12.0±4.4 HU) and then remained constantly low up to 200keV (11.6±4.3HU). For liver, pancreas, portal vein, aorta, muscle and fat, noise at all levels of VMI was lower than of conventional images (p<0.01). For all organs, SNR, and CNR were highest at 40keV (6.8-34.9; 18.3-44.9, respectively) after which they gradually decreased up to 120keV (3.4-6.5; 9.5-13.0) and then remained constant to 200keV (2.6-5.5; 8.5-12.5). Qualitative scores of VMI up to 70keV were significantly higher than the conventional images (p≤0.01), whereas for VMI≥80keV, they were lower than conventional images (p<0.001).

CONCLUSION: VMI obtained from the novel SDCT scanner have low noise across the entire spectrum of energies. There are significant SNR and CNR improvements compared to conventional 120 kVp images.

来源:Eur J Radiol,查看原文。版权归原作者所有。