文献速递介入腹部US

离体肝组织中高强度聚焦超声所致病灶的数值与实验评估

Numerical and Experimental Evaluation of High-Intensity Focused Ultrasound-Induced Lesions in Liver Tissue Ex Vivo.
PubMed 文献 · J Ultrasound Med · 2017-11-30

本研究通过数值模拟与离体猪/肝组织实验,评估凹面换能器在85 W和135 W功率下高强度聚焦超声所致加热与热剂量轮廓,结果显示实验与数值结果一致性良好,焦点平均温度偏差分别为12.8%和4.3%,为确定病灶大小和识别非线性效应提供参考。

要点

  • 高强度聚焦超声(HIFU)热消融的数值模拟与离体肝组织实验温度结果一致性良好。
  • 换能器功率85 W时焦点平均温度偏差为12.8%,135 W时为4.3%。
  • 研究提示HIFU操作中需考虑非线性声学效应以判断病灶大小。

核心结论

  • 数值模拟与离体肝组织实验的加热结果一致性良好,焦点平均温度偏差在85 W和135 W功率下分别为12.8%和4.3%,可为HIFU操作者判断病灶大小和识别非线性效应提供参考。

研究设计与样本

  • 研究采用数值模拟与离体肝组织实验对照,使用凹面换能器,工作功率为85 W和135 W,通过Westervelt方程与Khokhlov-Zabolotskaya-Kuznetsov方程模拟非线性声场,并测量两种功率下的热传递以与离体实验结果比较。

诊断标准与关键数值

  • 计算模拟与实验结果的热剂量轮廓用于评估消融区域;焦点平均温度偏差在换能器功率85 W时为12.8%,在135 W时为4.3%。

临床应用与阅片要点

  • 原文结论指出本研究为HIFU操作者确定病灶大小和识别HIFU过程中应考虑的非线性效应提供了指导。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Recent advances in the field of acoustics and piezoelectric and ultrasound transducers have led to new approaches to the diagnosis and treatment of certain diseases. One method of treatment with ultrasonic waves is high-intensity focused ultrasound (HIFU) treatment, which is a thermal therapeutic method used to treat malignant tumors. Although a variety of treatment-planning strategies using ultrasonic waves have been investigated, little clinical success has been achieved. Computational modeling is a powerful tool for predicting device performance.

METHODS: The heating induced by a concave transducer with operating powers of 85 and 135 W was studied, and the experimental results presented in this article verify its applicability. Numerical simulations of the nonlinear acoustic field were performed by using the Westervelt and Khokhlov-Zabolotskaya-Kuznetsov equations. Heat transfer was measured for the 2 operational powers, and the results were compared with ex vivo experimental results. In addition, thermal dose contours for both the simulation and experimental results were calculated to investigate the ablated area.

RESULTS: Good agreement was found between the experimental and numerical results. The results show that the average temperature deviations calculated at the focal point were 12.8% and 4.3% for transducer powers of 85 and 135 W, respectively.

CONCLUSIONS: This study provides guidance to HIFU practitioners in determining lesion size and identifying nonlinear effects that should be considered in HIFU procedures.

来源:J Ultrasound Med,查看原文。版权归原作者所有。