经离体人腹壁的大孔径成像评估
在离体实验中,通过三条离体尸体腹壁和水路对点靶和病灶体模成像,使用7.36厘米有效孔径在6.4厘米深度成像,并采用合成孔径技术测量不同孔径下的图像质量指标。结果显示,尽管与对照组相比有退化,横向分辨率随孔径增大而提高且对比度无损失;大孔径数据的空间复合使病灶可检测性大幅改善,对比噪声比相比大相干孔径提高83%-106%。研究提示需要开发用于高分辨率腹部诊断成像的大阵列。
要点
- 离体腹壁成像中,增大有效孔径可提高横向分辨率且不损失对比度。
- 大孔径数据的空间复合使病灶可检测性大幅改善,对比噪声比提高83%-106%。
- 研究提示需要开发用于高分辨率腹部诊断成像的大阵列。
核心结论
- 离体实验表明,即使存在腹壁引起的声学杂波和畸变,使用大有效孔径仍可实现高分辨率成像,空间复合可大幅改善病灶可检测性。
研究设计与样本
- 离体实验研究,使用点靶和病灶体模,经水路和三条离体尸体腹壁成像,采用7.36厘米有效孔径在6.4厘米深度成像,并通过合成孔径技术测量不同孔径下的图像质量指标。
关键影像表现/征象
- 在三种腹壁情况下,与对照组相比虽观察到退化,但横向分辨率随孔径增大而提高且对比度无损失。
- 大孔径数据的空间复合使病灶可检测性大幅改善,对比噪声比相比大相干孔径提高83%-106%。
诊断标准与关键数值
- 使用7.36厘米有效孔径在6.4厘米深度成像,图像质量指标在系列孔径大小下通过合成孔径技术测量。
- 空间复合后对比噪声比相比大相干孔径提高83%-106%。
临床应用与阅片要点
- 研究提示需要开发用于高分辨率腹部诊断成像的大阵列。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Current clinical abdominal imaging arrays are designed to maximize angular field of view rather than the extent of the coherent aperture. We illustrate, in ex vivo experiments, the use of a large effective aperture to perform high-resolution imaging, even in the presence of abdominal wall-induced acoustic clutter and aberration. Point and lesion phantom targets were imaged through a water path and through three excised cadaver abdominal walls to create different clinically relevant clutter effects with matched imaging targets. A 7.36-cm effective aperture was used to image the targets at a depth of 6.4 cm, and image quality metrics were measured over a range of aperture sizes using synthetic aperture techniques. In all three cases, although degradation compared with the control was observed, lateral resolution improved with increasing aperture size without loss of contrast. Spatial compounding of the large-aperture data drastically improved lesion detectability and produced contrast-to-noise ratio improvements of 83%-106% compared with the large coherent aperture. These studies indicate the need for the development of large arrays for high-resolution abdominal diagnostic imaging.
引用
- Bottenus N, Long W, Morgan M, et al. Evaluation of Large-Aperture Imaging Through the ex Vivo Human Abdominal Wall. Ultrasound Med Biol. 2017. PMID: 29249458. doi:10.1016/j.ultrasmedbio.2017.10.019
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。