文献速递心血管CT

第三代双源CT前瞻性心电触发冠状动脉CT血管造影中自适应探测器准直的评价

Evaluation of an adaptive detector collimation for prospectively ECG-triggered coronary CT angiography with third-generation dual-source CT.
PubMed 文献 · Eur Radiol · 2017-12-07

研究基于103例三重排除检查回顾性推导理想心脏扫描长度,在新扫描仪前瞻性心电触发序列模式下比较自适应准直与固定64×0.6 mm、96×0.6 mm准直的扫描长度准确性和剂量参数。自适应准直每次均能匹配理想扫描长度,固定64×0.6 mm和96×0.6 mm准直平均扫描长度分别延长15.4%和27.2%;自适应与64×0.6 mm准直的DLP接近(120 kV下83 vs. 89 mGycm),96×0.6 mm准直DLP高62.7%(135 mGycm),p < 0.001。

要点

  • 自适应探测器准直在前瞻性心电触发序列扫描中每次均能准确匹配理想心脏扫描长度。
  • 固定64×0.6 mm和96×0.6 mm准直平均扫描长度分别较理想长度延长15.4%和27.2%。
  • 120 kV下自适应与64×0.6 mm准直DLP接近(83 vs. 89 mGycm),96×0.6 mm准直DLP高62.7%(135 mGycm),p < 0.001。

核心结论

  • 自适应探测器准直用于前瞻性心电触发序列采集时,扫描长度调整精度可达到以往仅螺旋采集才能实现的水平,有助于在传统步进-曝光模式会导致患者剂量显著增加的情况下保持低曝光。

研究设计与样本

  • 研究基于103例三重排除检查回顾性推导人心脏理想扫描长度,输入新扫描仪以三种探测器设置行前瞻性心电触发序列心脏扫描:自适应准直、固定64×0.6 mm准直、固定96×0.6 mm准直,记录有效扫描长度、与理想扫描长度的偏差及剂量参数(CTDIvol、DLP)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the impact of an adaptive detector collimation on the dose parameters and accurateness of scan length adaption at prospectively ECG-triggered sequential cardiac CT with a wide-detector third-generation dual-source CT.

METHODS: Ideal scan lengths for human hearts were retrospectively derived from 103 triple-rule-out examinations. These measures were entered into the new scanner operated in prospectively ECG-triggered sequential cardiac scan mode with three different detector settings: (1) adaptive collimation, (2) fixed 64 × 0.6-mm collimation, and (3) fixed 96 × 0.6-mm collimation. Differences in effective scan length and deviation from the ideal scan length and dose parameters (CTDIvol, DLP) were documented.

RESULTS: The ideal cardiac scan length could be matched by the adaptive collimation in every case while the mean scanned length was longer by 15.4% with the 64 × 0.6 mm and by 27.2% with the fixed 96 × 0.6-mm collimation. While the DLP was almost identical between the adaptive and the 64 × 0.6-mm collimation (83 vs. 89 mGycm at 120 kV), it was 62.7% higher with the 96 × 0.6-mm collimation (135 mGycm), p < 0.001.

CONCLUSION: The adaptive detector collimation for prospectively ECG-triggered sequential acquisition allows for adjusting the scan length as accurate as this can only be achieved with a spiral acquisition. This technique allows keeping patient exposure low where patient dose would significantly increase with the traditional step-and-shoot mode.

KEY POINTS: • Adaptive detector collimation allows keeping patient exposure low in cardiac CT. • With novel detectors the desired scan length can be accurately matched. • Differences in detector settings may cause 62.7% of excessive dose.

来源:Eur Radiol,查看原文。版权归原作者所有。