采用自适应迭代重建算法在管电流降低75%条件下行冠状动脉钙化评分的准确性
连续163例患者行320排探测器CT 120 kVp两期扫描,低剂量期管电流为常规25%但扫描长度标准14 cm,常规剂量期用常规管电流但扫描长度缩短至10 cm。低剂量图像分别用滤波反投影(FBP)和三维自适应迭代剂量降低(IR)重建,常规剂量FBP为金标准。IR在剂量降低75%时Agatston评分准确,风险分类重分配仅4.9%,FBP为23.3%。
要点
- 低剂量IR重建与常规剂量FBP的Agatston评分无显著差异,而低剂量FBP评分显著升高。
- 低剂量FBP使23.3%患者心血管风险分类重分配,IR仅4.9%。
- 47例Agatston评分为零的患者中,低剂量FBP假阳性15例(31.9%),IR无假阳性。
核心结论
- 三维自适应迭代剂量降低技术可在辐射剂量降低75%的条件下维持准确的Agatston评分和风险分层。
研究设计与样本
- 连续163例患者行320排探测器CT 120 kVp两期扫描,低剂量期管电流为常规25%、扫描长度14 cm,常规剂量期用常规管电流、扫描长度10 cm;低剂量图像用FBP和三维自适应迭代剂量降低重建,常规剂量FBP为金标准。
诊断标准与关键数值
- 以常规剂量FBP的Agatston评分为金标准,低剂量FBP评分较金标准显著升高,低剂量IR与金标准无显著差异。
- 低剂量FBP使38例(23.3%)患者心血管风险分类重分配,低剂量IR仅8例(4.9%)。
- 47例Agatston评分为零的患者中,低剂量FBP假阳性15例(31.9%),低剂量IR无假阳性。
- 与常规剂量FBP相比,低剂量FBP图像噪声明显增加,低剂量IR噪声仍保持较低。
临床应用与阅片要点
- 低剂量条件下应使用迭代重建而非FBP进行冠状动脉钙化评分,以避免评分升高和风险分类错误。
- 对Agatston评分为零的患者,低剂量FBP可出现假阳性,迭代重建未见假阳性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To assess the accuracy of an iterative reconstruction (IR) technique for coronary artery calcium scoring with reduced radiation dose.
METHODS: 163 consecutive patients underwent twofold scanning by 320-row detector CT at 120 kVp. A low-dose scan at 25% tube current but with standard scan length (14 cm) was followed by a standard dose scan with routine tube current but reduced scan length (10 cm). Reduced dose images were constructed using filtered back-projection (FBP) and IR (adaptive iterative dose reduction in three dimensions). The standard dose scan reconstructed with FBP served as the gold standard for comparisons. Image noise and Agatston coronary calcium scores were determined and compared between the groups.
RESULTS: Compared with FBP at standard dose, noise at reduced dose increased markedly with FBP but remained low with IR. Mean Agatston score with FBP at reduced dose showed a significant increase as compared with FBP at standard dose. No significant difference was observed when applying IR at reduced dose. At reduced dose, 38 (23.3%) patients were reassigned to a different cardiovascular risk category with FBP but only 8 (4.9%) with IR. Out of 47 patients with a zero Agatston score, 15 patients (31.9%) were false-positive with FBP at reduced dose, but no false positives were found with IR.
CONCLUSION: IR allows accurate coronary artery calcium scoring with a radiation dose reduced by 75%. Advances in knowledge: The application of adaptive iterative dose reduction in three dimensions allows the maintenance of accurate Agatston scores and risk stratification at significantly reduced tube current, thus reducing the patient's exposure to ionizing radiation.