相位滤波移动集合相关方法用于超声心动图粒子图像测速的开发与验证

Development and Validation of a Phase-Filtered Moving Ensemble Correlation for Echocardiographic Particle Image Velocimetry.
PubMed 文献 · Ultrasound Med Biol · 2017-11-28

研究开发了采用移动集合相关与动态相位相关滤波的超声心动图粒子图像测速新方法,在计算生成图像中误差较标准相关法降低约一半,在8例舒张充盈受损患者中与彩色M型速度时间序列的一致性优于标准处理(R=0.90 vs 0.70),充盈传播速度和压力梯度与彩色M型的相关性也更高(R²=0.69 vs 0.15和0.07)。

要点

  • 移动集合相关法在计算生成图像中偏倚和随机误差均优于标准相关法(β=-0.06 vs -0.15;σ=0.49 vs 1.00)。
  • 8例舒张充盈受损患者中,移动集合相关法与彩色M型速度时间序列一致性R=0.90,标准法R=0.70。
  • 充盈传播速度与压力梯度在移动集合处理下与彩色M型相关R²=0.69,标准法仅R²=0.15和0.07。

核心结论

  • 新方法可提高超声心动图粒子图像测速的速度测量准确性,使常规临床扫描的处理更稳健,并可能提升EchoPIV评估左心室功能的实用性。

研究设计与样本

  • 方法开发与验证研究,使用计算生成的超声心动图图像进行误差分析,并在8例舒张充盈受损患者中与彩色M型速度时间序列对比验证。

关键影像表现/征象

  • 计算生成图像误差分析显示移动集合法偏倚β=-0.06,标准成对相关法β=-0.15。
  • 计算生成图像随机误差σ移动集合法=0.49,标准成对相关法σ=1.00。
  • 患者EchoPIV速度时间序列与彩色M型对比,移动集合法R=0.90,标准成对相关法R=0.70。
  • 充盈传播速度与彩色M型相关R²=0.69(移动集合法)vs R²=0.15(成对处理法)。
  • 一维室内压力梯度与彩色M型相关R²=0.69(移动集合法)vs R²=0.07(成对处理法)。

临床应用与阅片要点

  • 新方法采用移动集合相关与动态相位相关滤波,可提高EchoPIV速度测量准确性。
  • 在舒张充盈受损患者中,移动集合法获得的充盈传播速度和室内压力梯度与彩色M型相关性优于标准成对处理法。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

A new processing method for echocardiographic particle image velocimetry (EchoPIV) using moving ensemble (ME) correlation with dynamic phase correlation filtering was developed to improve velocity measurement accuracy for routine clinical evaluation of cardiac function. The proposed method was tested using computationally generated echocardiogram images. Error analysis indicated that ME EchoPIV yields a twofold improvement in bias and random error over the current standard correlation method (βPairwise = -0.15 vs. βME = -0.06; σPairwise = 1.00 vs. σME = 0.49). Subsequently a cohort of eight patients with impaired diastolic filling underwent similar evaluation. Comparison of patient EchoPIV velocity time series with corresponding color M-mode velocity time series revealed better agreement for ME EchoPIV compared with standard PIV processing (RME = 0.90 vs. RPairwise = 0.70). Further time series analysis was performed to measure filling propagation velocity and 1-D intraventricular pressure gradients. Comparison against CMM values indicated that both measurements are completely decorrelated for pairwise processing (R2Vp = 0.15, R2IVPD = 0.07), whereas ME processing correlates decently (R2Vp = 0.69, R2IVPD = 0.69). This new approach enables more robust processing of routine clinical scans and can increase the utility of EchoPIV for the assessment of left ventricular function.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Meyers BA, Goergen CJ, Vlachos PP, et al. Development and Validation of a Phase-Filtered Moving Ensemble Correlation for Echocardiographic Particle Image Velocimetry. Ultrasound Med Biol. 2017. PMID: 29195751. doi:10.1016/j.ultrasmedbio.2017.10.008

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。