基于胸部体层合成成像的计算机辅助检测系统对肺结节检出的价值
50例有肺结节和50例无肺结节患者同日行胸部体层合成和多排CT检查,15名观察者(实习生/住院医师、胸部放射科医师、腹部放射科医师各5名)盲法随机独立评估体层合成图像,先无CAD后有CAD标记,以多排CT为参考标准。汇总诊断效能使用CAD后显著优于不用CAD(FOM 0.74 vs 0.71,P=0.02),亚组分析显示实习生/住院医师使用CAD后诊断效能显著改善(FOM 0.70 vs 0.62,P=0.001),而胸部放射科医师和腹部放射科医师的FOM值虽有提高但无统计学显著性(分别为0.80 vs 0.78,P=0.38;0.74 vs 0.73,P=0.65)。
要点
- 汇总15名观察者,使用CAD后诊断效能FOM 0.74 vs 0.71,P=0.02,差异有统计学意义。
- 实习生和住院医师使用CAD后FOM 0.70 vs 0.62,P=0.001,诊断效能显著提升。
- 胸部放射科医师使用CAD后FOM 0.80 vs 0.78,P=0.38,差异无统计学意义。
核心结论
- CAD显著改善了实习生和住院医师在胸部体层合成图像上检出肺结节的诊断效能,但对胸部放射科医师和腹部放射科医师获益甚微。
研究设计与样本
- 研究纳入50例有肺结节和50例无肺结节患者,同日行胸部体层合成和多排CT检查,以多排CT图像为参考标准。
- 15名观察者(5名实习生和住院医师、5名胸部放射科医师、5名腹部放射科医师)以盲法随机方式独立评估100例患者的体层合成图像,先不用CAD,然后加入CAD标记。
- 统计分析采用自由响应受试者工作特征分析。
诊断标准与关键数值
- 汇总15名观察者的诊断效能使用CAD后FOM为0.74,不用CAD为0.71,P=0.02。
- 使用CAD时所有病例的平均真阳性率和假阳性率分别为0.56和0.26,不用CAD时分别为0.47和0.20。
- 实习生和住院医师使用CAD后FOM为0.70,不用CAD为0.62,P=0.001。
- 胸部放射科医师使用CAD后FOM为0.80,不用CAD为0.78,P=0.38。
- 腹部放射科医师使用CAD后FOM为0.74,不用CAD为0.73,P=0.65。
临床应用与阅片要点
- 原文结论指出CAD显著改善了实习生和住院医师在胸部体层合成图像上检出肺结节的诊断效能,但对胸部放射科医师和腹部放射科医师获益甚微。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To assess the value of a computer-aided detection (CAD) system for the detection of pulmonary nodules on chest tomosynthesis images. Materials and Methods Fifty patients with and 50 without pulmonary nodules underwent both chest tomosynthesis and multidetector computed tomography (CT) on the same day. Fifteen observers (five interns and residents, five chest radiologists, and five abdominal radiologists) independently evaluated tomosynthesis images of 100 patients for the presence of pulmonary nodules in a blinded and randomized manner, first without CAD, then with the inclusion of CAD marks. Multidetector CT images served as the reference standard. Free-response receiver operating characteristic analysis was used for the statistical analysis. Results The pooled diagnostic performance of 15 observers was significantly better with CAD than without CAD (figure of merit [FOM], 0.74 vs 0.71, respectively; P = .02). The average true-positive fraction and false-positive rate per all cases with CAD were 0.56 and 0.26, respectively, whereas those without CAD were 0.47 and 0.20, respectively. Subanalysis showed that the diagnostic performance of interns and residents was significantly better with CAD than without CAD (FOM, 0.70 vs 0.62, respectively; P = .001), whereas for chest radiologists and abdominal radiologists, the FOM with CAD values were greater but not significantly: 0.80 versus 0.78 (P = .38) and 0.74 versus 0.73 (P = .65), respectively. Conclusion CAD significantly improved diagnostic performance in the detection of pulmonary nodules on chest tomosynthesis images for interns and residents, but provided minimal benefit for chest radiologists and abdominal radiologists. © RSNA, 2017 Online supplemental material is available for this article.