头颈部MRI最佳脂肪抑制:多点Dixon与两种不同脂肪抑制技术(频谱预饱和反转恢复和STIR)的比较
回顾性分析72例连续患者的头颈部MRI数据,按脂肪抑制技术分为A组(STIR及频谱预饱和反转恢复钆T1WI)和B组(多点Dixon T2 TSE及多点Dixon钆T1WI TSE),最终纳入64例(A组33例、B组31例)。多点Dixon T2和钆T1WI技术的信号强度比、整体图像质量和脂肪抑制分级均显著高于STIR和频谱预饱和反转恢复钆T1WI,且扫描时间更短。结论认为多点Dixon技术相比STIR和频谱预饱和反转恢复钆T1WI技术提供更好的图像质量和均匀的脂肪抑制,扫描时间更短。
要点
- 多点Dixon T2和钆T1WI的信号强度比显著高于STIR(P < .001)和频谱预饱和反转恢复钆T1WI(P < .001)。
- 多点Dixon T2和钆T1WI的整体图像质量及脂肪抑制分级显著优于STIR和频谱预饱和反转恢复钆T1WI。
- 多点Dixon扫描时间较短:T2序列2分钟 vs STIR 4分56秒;钆T1WI序列2分钟 vs 频谱预饱和反转恢复3分钟。
核心结论
- 多点Dixon技术相比STIR和频谱预饱和反转恢复钆T1WI技术提供更好的图像质量和均匀的脂肪抑制,且扫描时间更短。
研究设计与样本
- 回顾性研究,连续72例患者头颈部MRI数据集,按脂肪抑制技术分为A组(STIR和频谱预饱和反转恢复钆T1WI)和B组(多点Dixon T2 TSE和多点Dixon钆T1WI TSE),最终纳入64例(A组33例、B组31例),采用Mann-Whitney U检验比较客观和主观图像质量及扫描时间。
关键影像表现/征象
- 多点Dixon T2和钆T1WI技术的信号强度比分别显著高于STIR(P < .001)和频谱预饱和反转恢复钆T1WI(P < .001)。
- 两名独立盲法阅片者评估显示,多点Dixon T2和钆T1WI技术的整体图像质量分别显著高于STIR(P = .022)和频谱预饱和反转恢复钆T1WI(P < .001)。
- 多点Dixon T2和钆T1WI技术在3个不同区域的脂肪抑制分级分别显著高于STIR(P < .013)和频谱预饱和反转恢复钆T1WI(P < .001)。
诊断标准与关键数值
- 信号强度比:多点Dixon T2显著高于STIR(P < .001),多点Dixon钆T1WI显著高于频谱预饱和反转恢复钆T1WI(P < .001)。
- 整体图像质量评分:多点Dixon T2显著高于STIR(P = .022),多点Dixon钆T1WI显著高于频谱预饱和反转恢复钆T1WI(P < .001)。
- 脂肪抑制分级(3个不同区域):多点Dixon T2显著高于STIR(P < .013),多点Dixon钆T1WI显著高于频谱预饱和反转恢复钆T1WI(P < .001)。
- 扫描采集时间:T2加权序列多点Dixon 2分钟 vs STIR 4分56秒;钆T1WI序列多点Dixon 2分钟 vs 频谱预饱和反转恢复3分钟。
临床应用与阅片要点
- 多点Dixon技术可提供更好的图像质量和均匀脂肪抑制,且扫描时间较短,适用于头颈部脂肪抑制MRI。
- 原文比较了多点Dixon T2 TSE与STIR、多点Dixon钆T1WI TSE与频谱预饱和反转恢复钆T1WI,阅片时需注意不同序列的脂肪抑制分级差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Uniform complete fat suppression is essential for identification and characterization of most head and pathology. Our aim was to compare the multipoint Dixon turbo spin-echo fat-suppression technique with 2 different fat-suppression techniques, including a hybrid spectral presaturation with inversion recovery technique and an inversion recovery STIR technique, in head and neck fat-suppression MR imaging.
MATERIALS AND METHODS: Head and neck MR imaging datasets of 72 consecutive patients were retrospectively reviewed. All patients were divided into 2 groups based on the type of fat-suppression techniques used (group A: STIR and spectral presaturation with inversion recovery gadolinium-T1WI; group B: multipoint Dixon T2 TSE and multipoint Dixon gadolinium-T1WI TSE). Objective and subjective image quality and scan acquisition times were assessed and compared between multipoint Dixon T2 TSE versus STIR and multipoint Dixon gadolinium-T1WI TSE versus spectral presaturation with inversion recovery gadolinium-T1WI using the Mann-Whitney U test.
RESULTS: A total of 64 patients were enrolled in the study (group A, n = 33 and group B, n = 31). Signal intensity ratios were significantly higher for multipoint Dixon T2 and gadolinium-T1WI techniques compared with STIR (P < .001) and spectral presaturation with inversion recovery gadolinium-T1WI (P < .001), respectively. Two independent blinded readers revealed that multipoint Dixon T2 and gadolinium-T1WI techniques had significantly higher overall image quality (P = .022 and P < .001) and fat-suppression grades (P < .013 and P < .001 across 3 different regions) than STIR and spectral presaturation with inversion recovery gadolinium-T1WI, respectively. The scan acquisition time was relatively short for the multipoint Dixon technique (2 minutes versus 4 minutes 56 seconds for the T2-weighted sequence and 2 minutes versus 3 minutes for the gadolinium-T1WI sequence).
CONCLUSIONS: The multipoint Dixon technique offers better image quality and uniform fat suppression at a shorter scan time compared with STIR and spectral presaturation with inversion recovery gadolinium-T1WI techniques.