婴儿脑部成像中静音T2 PROPELLER序列的图像质量评估
研究在3 T MR扫描仪上对24名婴儿共36次扫描分别采集常规与静音T2 PROPELLER序列,静音序列声压级73.3 dB较常规92.1 dB降低20%,两序列在相对对比度、信噪比及多项定性图像质量指标上均无显著差异。
要点
- 静音T2 PROPELLER序列声压级73.3 dB,较常规序列92.1 dB降低20%。
- 两序列相对对比度(0.069 vs 0.068)与信噪比(75.4 vs 114.8)无显著差异。
- 整体图像质量、灰白质分化、分流相关伪影、运动伪影及髓鞘化程度均无显著差异。
核心结论
- 静音T2 PROPELLER序列可降低声学噪声,并产生与常规序列相当的图像质量。
研究设计与样本
- 对24名婴儿共36次扫描,使用3 T MR扫描仪,每名患者均行常规与静音T2 PROPELLER序列,测量声学噪声水平并进行定量与定性图像评估。
诊断标准与关键数值
- 常规T2 PROPELLER序列声压级92.1 dB,静音T2 PROPELLER序列声压级73.3 dB,降低20%。
- 定量评估:常规 vs 静音T2 PROPELLER相对对比度0.069 vs 0.068(p值=0.536),信噪比75.4 vs 114.8(p值=0.098),均无显著差异。
- 定性评估:整体图像质量(p值=0.572)、灰白质分化(p值=0.986)、分流相关伪影(p值>0.999)、运动伪影(p值=0.801)及不同脑区髓鞘化程度(p值≥0.092)均无显著差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Infants are vulnerable to high acoustic noise. Acoustic noise generated by MR scanning can be reduced by a silent sequence. The purpose of this study is to compare the image quality of the conventional and silent T2 PROPELLER sequences for brain imaging in infants.
METHODS: A total of 36 scans were acquired from 24 infants using a 3 T MR scanner. Each patient underwent both conventional and silent T2 PROPELLER sequences. Acoustic noise level was measured. Quantitative and qualitative assessments were performed with the images taken with each sequence.
RESULTS: The sound pressure level of the conventional T2 PROPELLER imaging sequence was 92.1 dB and that of the silent T2 PROPELLER imaging sequence was 73.3 dB (reduction of 20%). On quantitative assessment, the two sequences (conventional vs silent T2 PROPELLER) did not show significant difference in relative contrast (0.069 vs 0.068, p value = 0.536) and signal-to-noise ratio (75.4 vs 114.8, p value = 0.098). Qualitative assessment of overall image quality (p value = 0.572), grey-white differentiation (p value = 0.986), shunt-related artefact (p value > 0.999), motion artefact (p value = 0.801) and myelination degree in different brain regions (p values ≥ 0.092) did not show significant difference between the two sequences.
CONCLUSION: The silent T2 PROPELLER sequence reduces acoustic noise and generated comparable image quality to that of the conventional sequence. Advances in knowledge: This is the first report to compare silent T2 PROPELLER images with that of conventional T2 PROPELLER images in children.