前列腺近距离放射治疗超声引导系统质量保证推荐标准的综述
本文综述了各标准机构与专业组织针对前列腺近距离放射治疗中超声引导系统的质量保证测试指南与测试体模,指出当前推荐测试并非针对该临床应用,而是通用性的,且缺乏临床相关的测试体模;文章还批判性分析了最能反映超声引导前列腺近距离放射治疗中成像与引导任务的测试,以期为设计经直肠超声质量保证协议提供依据。
要点
- 现有前列腺近距离放射治疗超声系统的推荐测试为通用性测试,与腹部扫描等应用类似,并非针对该临床操作。
- 目前没有针对前列腺近距离放射治疗超声系统质量保证的临床相关测试体模推荐。
- 需监测该应用中的图像质量以确保操作信心的一致性,并应设计经直肠超声质量保证协议。
核心结论
- 当前推荐标准中用于前列腺近距离放射治疗超声系统的测试似乎并非针对该临床应用,而是通用性的;且缺乏临床相关测试体模,需设计经直肠超声质量保证协议。
研究设计与样本
- 综述,回顾不同标准机构与专业组织针对前列腺近距离放射治疗超声系统的推荐测试指南与测试体模,并对最能反映该操作中成像与引导任务的测试进行批判性分析。
临床应用与阅片要点
- 作者指出应设计针对前列腺近距离放射治疗中成像与引导任务的经直肠超声质量保证协议,并批判性分析最能反映这些任务的测试。
- 作者指出图像质量需被监测以确保对该操作信心的一致性。
局限性
- 推荐测试似乎并非针对超声引导前列腺近距离放射治疗的临床应用,而是通用性的,与其他临床应用(如腹部扫描)的推荐类似。
- 目前没有临床相关的测试体模被推荐用于前列腺近距离放射治疗超声系统的质量保证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Ultrasound guided brachytherapy for the treatment of prostate cancer has become a routine treatment option, due to many benefits including patient recovery and dose localisation [1]; however it is not clear whether the standards which govern the image quality for these systems are adequate. Upon review of the recommended standards for ultrasound systems used in prostate brachytherapy procedures, the recommended tests do not appear to be specific to the clinical application of ultrasound guided prostate brachytherapy. Rather they are generic and similar to those recommended for other clinical applications such as general abdominal scanning [2]. Furthermore, there is growing evidence that these tests should be specific to the clinical application [3,4] in order to gain meaningful data about the performance of the system for the application, and also to detect clinically relevant changes in quality control results. An additional problem is that there are no clinically relevant test phantom recommended for the quality assurance of ultrasound systems used in prostate brachytherapy. The image quality for this application of ultrasound needs to be monitored to ensure consistent levels of confidence in the procedure. This paper reviews the currently recommended test guidelines and test phantoms for ultrasound systems used in prostate brachytherapy from the different standard bodies and professional organisations. A critical analysis of those tests which are most reflective of the imaging and guidance tasks undertaken in an ultrasound guided prostate brachytherapy procedure will also be presented to inform the design of a TRUS quality assurance protocol.
引用
- Doyle AJ, King DM, Browne JE, et al. A review of the recommendations governing quality assurance of ultrasound systems used for guidance in prostate brachytherapy. Phys Med. 2017. PMID: 29254591. doi:10.1016/j.ejmp.2017.11.011
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。