文献速递胸部CT

亚毫西弗胸部双能CT:一项先导研究

Submillisievert chest dual energy computed tomography: a pilot study.
PubMed 文献 · Br J Radiol · 2017-12-05

一项经IRB批准的前瞻性先导研究纳入20例患者,在标准胸部双能CT扫描后10秒内追加80/140 kV低剂量序列,评估亚毫西弗剂量下图像质量与诊断信息。结果显示图像噪声增加但整体质量和锐利度均为良好或优秀(20/20),有效剂量0.7±0.1 mSv,作者认为常规胸部双能CT可在亚毫西弗剂量下完成且不丢失相关诊断信息。

要点

  • 低剂量序列有效剂量0.7±0.1 mSv,CTDIvol 1.3±0.2 mGy,DLP 51±9.9 mGy·cm。
  • 图像噪声增加但整体质量和锐利度良好或优秀,所有病例(20/20)均满足诊断。
  • 观察者内和观察者间一致性κ分别为0.83和0.73。

核心结论

  • 常规胸部双能CT可在亚毫西弗剂量下完成,图像质量良好且不丢失相关诊断信息;作者称这是首次在患者研究中用双源双能CT扫描仪成功实现亚毫西弗剂量。

研究设计与样本

  • IRB批准的前瞻性先导研究,纳入20例患者,在双源多排CT(80/140 kV)上于标准扫描后10秒内追加低剂量序列,通过降低质量参考毫安秒并结合角度曝光控制实现剂量降低;4名对临床资料设盲的读者评估图像。

关键影像表现/征象

  • 低剂量图像噪声增加,但融合图像整体质量和锐利度在全部病例(20/20)中评为良好或优秀。
  • 低剂量图像上所有发现均边界清晰。

诊断标准与关键数值

  • 观察者内一致性κ=0.83,观察者间一致性κ=0.73。
  • 低剂量双能CT平均CTDIvol 1.3±0.2 mGy、体型特异性剂量估计1.8±0.2 mGy、DLP 51±9.9 mGy·cm、有效剂量0.7±0.1 mSv。

临床应用与阅片要点

  • 作者认为常规胸部双能CT可在亚毫西弗剂量下进行,图像质量良好且不丢失相关诊断信息。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To assess if diagnostic dual energy CT (DECT) of the chest can be achieved at submillisievert (sub-mSv) doses.

METHODS: Our IRB-approved prospective study included 20 patients who were scanned on dual-source multidector CT(MDCT). All patients gave written informed consent for acquisition of additional image series at reduced radiation dose on a dual-source MDCT (80/140 kV) within 10 s after the standard of care acquisition. Dose reduction was achieved by reducing the quality reference milliampere-second, with combined angular exposure control. Four readers, blinded to all clinical data, evaluated the image sets. Image noise, signal-to-noise and contrast-to-noise ratio were assessed. Volumetric CT dose index (CTDIvol), doselength product (DLP), size specific dose estimate, and effective dose were also recorded.

RESULTS: The mean age and body mass index of the patients were 71 years ± 9 and 24 kg m-2 ± 3, respectively. Although images became noisier, overall image quality and image sharpness on blended images were considered good or excellent in all cases (20/20). All findings made on the reduced dose images presented with good demarcation. The intraobserver and interobserver agreements were κ = 0.83 and 0.73, respectively. Mean CTDIvol, size specific dose estimate, DLP and effective dose for reduced dose DECT were: 1.3 ± 0.2 mGy, 1.8 ± 0.2 mGy, 51 ± 9.9 mGy.cm and 0.7 ± 0.1 mSv, respectively.

CONCLUSION: Routine chest DECT can be performed at sub-mSv doses with good image quality and without loss of relevant diagnostic information. Advances in knowledge: (1) Contrast-enhanced DECT of the chest can be performed at sub-mSv doses, down to mean CTDIvol 1.3 mGy and DLP 51 mGy.cm in patients with body mass index <31 kg m-2. (2) To our knowledge, this is the first time that sub-mSv doses have been successfully applied in a patient study using a dual source DECT scanner.

来源:Br J Radiol,查看原文。版权归原作者所有。