降噪血管造影成像技术降低支气管动脉栓塞术中的辐射剂量

Noise reduction angiographic imaging technology reduces radiation dose during bronchial artery embolization.
PubMed 文献 · Eur J Radiol · 2017-10-31

回顾性研究70例支气管动脉栓塞术患者,比较技术升级前(32例)与升级后(38例)的辐射剂量。升级后累积剂量面积乘积降低59%、累积空气比释动能降低60%,曝光帧数、透视时间和对比剂用量无显著差异,图像质量无显著差异。

要点

  • 降噪成像技术使剂量面积乘积降低59%、空气比释动能降低60%。
  • 曝光帧数、透视时间和对比剂用量在升级前后无显著差异。
  • 两名盲法阅片者评估图像质量在升级前后无显著差异。

核心结论

  • 新型血管造影降噪技术在不损害图像质量、不增加透视时间或对比剂用量的情况下,显著降低支气管动脉栓塞术中的辐射剂量。

研究设计与样本

  • 回顾性研究,共70例接受支气管动脉栓塞术的患者,技术升级前32例、升级后38例;升级后技术包含附加滤过和改进的图像处理;记录累积空气比释动能、累积剂量面积乘积、曝光帧数、总透视时间和对比剂用量,采用双尾t检验比较;数字减影血管造影图像质量由两名盲法阅片者独立评估,采用Wilcoxon符号秩检验比较。

诊断标准与关键数值

  • 累积剂量面积乘积:升级前149.2(103.1-279.1)Gy*cm2 vs 升级后54.8(38.2-100.7)Gy*cm2,升级后降低59%,p<0.001。
  • 累积空气比释动能:升级前1.3(0.6-1.9)Gy vs 升级后0.5(0.3-0.9)Gy,升级后降低60%,p<0.001。
  • 曝光帧数:升级前251±181帧 vs 升级后254±133帧,p=0.07,无显著差异。
  • 透视时间:升级前28.8(18.5-50.4)min vs 升级后28.1(23.3-38.7)min,p=0.73,无显著差异。
  • 对比剂用量:升级前139.5±70.8 ml vs 升级后163.1±63.1 ml,p=0.11,无显著差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Comparison of radiation doses in patients undergoing angiographic bronchial artery embolization (BAE) before and after a noise reduction imaging technology upgrade.

METHODS: We performed a retrospective study of 70 patients undergoing BAE. Procedures were performed before (n=32) and after (n=38) the technology upgrade containing additional filters and improved image-processing. Cumulative air kerma (AK), cumulative dose area product (DAP), number of exposure frames, total fluoroscopy time and amount of contrast agent were recorded. Mean values were calculated and compared using two-tailed t-tests. DSA image quality was assessed independently by two blinded readers and compared using the Wilcoxon signed-rank test.

RESULTS: Using the new technology resulted in a significant reduction of 59% in DAP (149.2 (103.1-279.1) vs. 54.8 (38.2-100.7) Gy*cm2, p<0.001) and a significant reduction of 60% for AK (1.3 (0.6-1.9) vs. 0.5 (0.3-0.9) Gy, p<0.001) in comparison to procedures before the upgrade. There was no significant difference between the number of exposure frames in both groups (251±181 vs. 254±133 frames, p=0.07), time of fluoroscopy (28.8 (18.5-50.4) vs. 28.1 (23.3-38.7) min, p=0.73), or the amount of contrast agent used (139.5±70.8 vs. 163.1±63.1ml, p=0.11). No significant difference regarding image quality could be detected (3 (2,3) vs. 3 (2-4), p=0.64).

CONCLUSIONS: The new angiographic noise reduction technology significantly decreases the radiation dose during bronchial artery embolization without compromising image quality or increasing time of fluoroscopy or contrast volume.

来源:Eur J Radiol,查看原文。版权归原作者所有。