年龄相关骨电子密度对儿童MRI-only放疗计划中千伏与兆伏光子及电子线剂量分布计算的影响
本研究计算15岁以下儿童骨含量并代入橡树岭国家实验室儿童体模,用Xcom与MCNPX评估150 kVp、6 MV、18 MV光子及6 MeV电子束的衰减与剂量差异。结果显示150 kVp下11岁与新生儿衰减系数差近5%、成人与新生儿约8%;6 MeV电子在1岁与10岁间百分深度剂量差18.5%、10岁与15岁间差24%。作者认为MRI-only儿童放疗计划应按年龄组分别定义骨密度以准确计算剂量。
要点
- 150 kVp下11岁与新生儿衰减系数差近5%,成人与新生儿约8%。
- 6 MeV电子百分深度剂量1岁与10岁差18.5%,10岁与15岁差24%。
- 作者认为儿童MRI-only放疗计划需按年龄组分别定义骨密度。
核心结论
- 作者认为,对于光子或电子束放疗的MRI-only治疗计划,应按每个年龄组分别定义骨密度以进行准确剂量计算。
研究设计与样本
- 计算15岁以下儿童骨含量并代入橡树岭国家实验室儿童体模,使用Xcom与MCNPX蒙特卡洛程序计算150 kVp与6 MV光子及6 MeV电子束的百分深度剂量和束流分布。
诊断标准与关键数值
- 150 kVp X射线:11岁儿童与新生儿衰减系数差异近5%,成人与新生儿差异约8%。
- 18 MV光子束:11岁儿童与新生儿辐射衰减差异4%,成人与新生儿差异约7.4%。
- 6 MeV电子束:剂量差异出现在至2 cm深度处;1岁与10岁儿童百分深度剂量差异18.5%,10岁与15岁儿童差异24%。
临床应用与阅片要点
- 作者认为,儿童MRI-only光子或电子束放疗计划应按年龄组分别定义骨密度以实现准确剂量计算。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: MRI-only treatment planning (TP) can be advantageous in paediatric radiotherapy. However, electron density extraction is necessary for dose calculation. Normally, after bone segmentation, a bulk density is assigned. However, the variation of bone bulk density in patients makes the creation of pseudo CTs challenging. This study aims to assess the effects of bone density variations in children on radiation attenuation and dose calculation for MRI-only TP.
METHODS: Bone contents of <15-year-old children were calculated, and substituted in the Oak Ridge National Laboratory paediatric phantoms. The percentage depth dose and beam profile of 150 kVp and 6 MV photon and 6 MeV electron beams were then calculated using Xcom, MCNPX (Monte Carlo N-particle version X) and ORLN phantoms.
RESULTS: Using 150 kVp X-rays, the difference in attenuation coefficient was almost 5% between an 11-year-old child and a newborn, and ~8% between an adult and a newborn. With megavoltage radiation, the differences were smaller but still important. For an 18 MV photon beam, the difference of radiation attenuation between an 11-year-old child and a newborn was 4% and ~7.4% between an adult and a newborn. For 6 MeV electrons, dose differences were observed up to the 2 cm depth. The percentage depth dose difference between 1 and 10-year-olds was 18.5%, and between 10 and 15-year-olds was 24%.
CONCLUSION: The results suggest that for MRI-only TP of photon- or electron-beam radiotherapy, the bone densities of each age group should be defined separately for accurate dose calculation. Advances in knowledge: This study highlights the need for more age-specific determination of bone electron density for accurate dose calculations in paediatric MRI-only radiotherapy TP.
引用
- Zeinali-Rafsanjani B, Faghihi R, Mosleh-Shirazi MA, et al. Effect of age-dependent bone electron density on the calculated dose distribution from kilovoltage and megavoltage photon and electron radiotherapy in paediatric MRI-only treatment planning. Br J Radiol. 2017. PMID: 29091480. doi:10.1259/bjr.20170511
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。