定量血管壁T1 mapping对颈动脉斑块内出血的成像:技术开发与初步经验

Carotid Intraplaque Hemorrhage Imaging with Quantitative Vessel Wall T1 Mapping: Technical Development and Initial Experience.
PubMed 文献 · Radiology · 2017-11-08

本研究开发了基于三维黄金角径向k空间采样的GOAL-SNAP序列用于定量血管壁T1 mapping,体模实验中与二维反转恢复自旋回波序列T1测量高度相关(R2=0.99),在6名健康志愿者和5名动脉粥样硬化患者中,斑块内出血层面的最小T1显著低于无出血层面(371毫秒±93 vs 944毫秒±120,P=.01)。

要点

  • GOAL-SNAP序列可在5分钟内完成0.8毫米各向同性高分辨率血管壁T1 mapping。
  • 斑块内出血层面最小T1显著低于无出血层面(371毫秒±93 vs 944毫秒±120,P=.01)。
  • 正常血管壁和肌肉的估算T1分别为1195毫秒±136和1117毫秒±153。

核心结论

  • GOAL-SNAP序列实现了高空间分辨率(0.8毫米各向同性)且省时(5分钟)的血管壁T1 mapping,该序列可能提供更具定量可重复性的生物标志物来表征斑块内出血并监测其进展。

研究设计与样本

  • 技术开发与初步验证研究;体模实验比较GOAL-SNAP与二维反转恢复自旋回波序列的T1测量准确性;体内研究纳入6名健康志愿者(平均年龄27.8岁±3.0,年龄范围24-32岁,5名男性)和5名动脉粥样硬化患者(平均年龄66.4岁±5.5,范围60-73岁,5名男性),比较有和无斑块内出血的血管壁层面(每条动脉5个层面)的T1测量值;统计方法包括Pearson相关系数、Bland-Altman分析和按受试者置换的Wilcoxon秩和检验。

关键影像表现/征象

  • 斑块内出血层面的最小T1显著低于无斑块内出血层面(均值371毫秒±93 vs 944毫秒±120,P=.01)。
  • 正常血管壁的估算T1为1195毫秒±136,肌肉的估算T1为1117毫秒±153。

诊断标准与关键数值

  • 体模T1测量中GOAL-SNAP与反转恢复自旋回波序列高度相关(R2=0.99),平均偏差-25.8毫秒±43.6,平均百分比误差4.3%±2.5。
  • 斑块内出血层面与无斑块内出血层面的最小T1差异有统计学意义(371毫秒±93 vs 944毫秒±120,P=.01)。

临床应用与阅片要点

  • GOAL-SNAP序列可在5分钟内提供0.8毫米各向同性的血管壁T1定量图,可用于斑块内出血的定量评估。
  • 斑块内出血层面表现为最小T1值降低(371毫秒±93),无斑块内出血层面最小T1为944毫秒±120,阅片时可关注T1低值区域。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To develop a three-dimensional (3D) high-spatial-resolution time-efficient sequence for use in quantitative vessel wall T1 mapping. Materials and Methods A previously described sequence, simultaneous noncontrast angiography and intraplaque hemorrhage (SNAP) imaging, was extended by introducing 3D golden angle radial k-space sampling (GOAL-SNAP). Sliding window reconstruction was adopted to reconstruct images at different inversion delay times (different T1 contrasts) for voxelwise T1 fitting. Phantom studies were performed to test the accuracy of T1 mapping with GOAL-SNAP against a two-dimensional inversion recovery (IR) spin-echo (SE) sequence. In vivo studies were performed in six healthy volunteers (mean age, 27.8 years ± 3.0 [standard deviation]; age range, 24-32 years; five male) and five patients with atherosclerosis (mean age, 66.4 years ± 5.5; range, 60-73 years; five male) to compare T1 measurements between vessel wall sections (five per artery) with and without intraplaque hemorrhage (IPH). Statistical analyses included Pearson correlation coefficient, Bland-Altman analysis, and Wilcoxon rank-sum test with data permutation by subject. Results Phantom T1 measurements with GOAL-SNAP and IR SE sequences showed excellent correlation (R2 = 0.99), with a mean bias of -25.8 msec ± 43.6 and a mean percentage error of 4.3% ± 2.5. Minimum T1 was significantly different between sections with IPH and those without it (mean, 371 msec ± 93 vs 944 msec ± 120; P = .01). Estimated T1 of normal vessel wall and muscle were 1195 msec ± 136 and 1117 msec ± 153, respectively. Conclusion High-spatial-resolution (0.8 mm isotropic) time-efficient (5 minutes) vessel wall T1 mapping is achieved by using the GOAL-SNAP sequence. This sequence may yield more quantitative reproducible biomarkers with which to characterize IPH and monitor its progression. © RSNA, 2017.

来源:Radiology,查看原文。版权归原作者所有。