手臂牵引装置对颈部CT图像质量和辐射暴露的影响:一项前瞻性研究

Effect of an Arm Traction Device on Image Quality and Radiation Exposure during Neck CT: A Prospective Study.
PubMed 文献 · AJNR Am J Neuroradiol · 2017-11-09

前瞻性纳入73例完全缓解的淋巴瘤患者,在同一CT系统上先后行手臂牵引装置干预方案与标准方案颈部CT复查,平均间隔155天。干预方案使下颈部噪声降低25.2%-28.5%(P < .001),条纹伪影显著减少(P < .001),剂量长度乘积413 vs 397(P < .001),容积CT剂量指数13.9 vs 13.4(P < .001),双肩线平均下移2.1 cm。

要点

  • 手臂牵引装置使下颈部图像噪声降低25.2%-28.5%(P < .001)。
  • 条纹伪影显著减少(P < .001),双肩线平均下移2.1 cm。
  • 剂量长度乘积413 vs 397、容积CT剂量指数13.9 vs 13.4,均P < .001。

核心结论

  • 手臂牵引装置可改善颈部CT图像质量并降低辐射暴露,该装置可简单应用于配合良好、怀疑下颈部病变的患者,较常规成像方案具有明显优势。

研究设计与样本

  • 前瞻性研究,纳入73例完全缓解的淋巴瘤患者,在同一CT系统上先后行手臂牵引装置干预方案与无体位优化的标准方案两次连续颈部CT复查,平均间隔155天。

关键影像表现/征象

  • 干预方案下颈部平均噪声降低25.2%-28.5%(P < .001)。
  • 干预方案显著减少锁骨上窝条纹伪影(P < .001)。
  • 干预方案使双肩线平均下移2.1 cm。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: The image quality of neck CT is frequently disturbed by streak artifact from the shoulder girdles. Our aim was to determine the effects of an arm traction device on image quality and radiation exposure in neck CT.

MATERIALS AND METHODS: Patients with lymphoma with complete remission who were scheduled to undergo 2 consecutive follow-up neck CT scans for surveillance within a 1-year interval were enrolled in this prospective study. They underwent 2 consecutive neck CT scans (intervention protocol: patients with an arm traction device; standard protocol: no positioning optimization) on the same CT system. The primary outcome measures were image noise in the lower neck and dose-length product. Secondary outcomes were streak artifacts in the supraclavicular fossa, volume CT dose index, and the extent of the biacromial line shift.

RESULTS: Seventy-three patients were enrolled and underwent 2 consecutive CT scans with a mean interval of 155 days. In the intervention protocol, a mean noise reduction in the lower neck of 25.2%-28.5% (P < .001) was achieved, and a significant decrease in dose-length product (413 versus 397, P < .001) was observed. The intervention protocol significantly decreased streak artifacts (P < .001) and volume CT dose index (13.9 versus 13.4, P < .001) and could lower the biacromial line an average of 2.1 cm.

CONCLUSIONS: An arm traction device can improve image quality and reduce radiation exposure during neck CT. The device can be simply applied in cooperative patients with suspected lower neck lesions, and the approach offers distinct advantages over the conventional imaging protocol.