乳腺X线摄影中乳腺动脉钙化的半自动定量工具:观察者内与观察者间可重复性研究

Breast arterial calcifications on mammography: intra- and inter-observer reproducibility of a semi-automatic quantification tool.
PubMed 文献 · Radiol Med · 2017-10-30

本研究回顾性选取53例受试者的212幅乳腺X线影像(106幅头尾位、106幅内外斜位),由两名经验超过6个月的放射诊断住院医师使用半自动软件独立两次勾画乳腺动脉钙化感兴趣区,评估观察者内与观察者间可重复性。结果显示观察者内偏倚11.9 mm²、可重复性系数32.7 mm²、平均测量值72.8 mm²、可重复性55%;观察者间对应数据为-7.0、61.4、63.4 mm²和仅3%。作者结论认为该半自动工具可重复性差,钙化的人工识别是变异的主要来源,需进一步研究才能将钙化定量转化为临床实践。

要点

  • 半自动定量工具观察者间可重复性仅3%,观察者内为55%,整体可重复性差。
  • 观察者间偏倚-7.0 mm²、可重复性系数61.4 mm²,平均测量值63.4 mm²。
  • 作者认为乳腺动脉钙化的人工识别是变异的主要来源,需进一步研究才能用于临床实践。

核心结论

  • 该半自动乳腺动脉钙化定量工具可重复性差,人工识别钙化是变异的主要来源,需进一步研究才能将钙化定量转化为临床实践。

研究设计与样本

  • 回顾性研究,从53例受试者中选取212幅乳腺X线影像(106幅头尾位、106幅内外斜位),由两名接受过强化培训、乳腺X线摄影经验超过6个月的放射诊断住院医师使用半自动软件独立两次分割图像。

诊断标准与关键数值

  • 观察者1总钙化量为56.6(IQR 18.1-91.1),观察者2为41.0(IQR 18.8-90.9)。观察者内Bland-Altman分析:偏倚11.9 mm²,可重复性系数32.7 mm²,平均测量值72.8 mm²,可重复性55%。观察者间Bland-Altman分析:偏倚-7.0 mm²,可重复性系数61.4 mm²,平均测量值63.4 mm²,可重复性3%。

局限性

  • 该半自动工具可重复性差。
  • 人工识别乳腺动脉钙化是变异的主要来源。
  • 需进一步研究才能将乳腺动脉钙化定量转化为临床实践。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: A strong association between breast arterial calcifications (BAC) and cardiovascular disease has been demonstrated. However, BAC quantification tools are lacking. We evaluated the intra- and inter-observer reproducibility of a semi-automatic tool for BAC quantification on digital mammograms.

MATERIALS AND METHODS: A multivendor image dataset of 212 mammographic views, 106 cranio-caudal (CC) and 106 medio-lateral oblique (MLO), were retrospectively selected from 53 subjects if BAC were seen in at least one view. Images were segmented twice by two intensively trained residents in Radiodiagnostics with > 6-month experience in mammography using a semi-automatic software. The two observers (O1, O2) independently positioned rectangular ROIs where they recognized BAC on both CC and MLO views, separately. The adaptive thresholding algorithm automatically provided the BAC amount in mm2. Number, size, and position of the ROIs were observer-dependent. Total BAC amount was calculated for each patient. Bland-Altman analysis was used.

RESULTS: Total BAC amount was 56.6 (IQR 18.1-91.1) and 41.0 (IQR 18.8-90.9) for O1 and O2, respectively. Intra-observer Bland-Altman analysis showed a bias of 11.9 mm2, a coefficient of repeatability of 32.7 mm2, an average measurement of 72.8 mm2, for a 55% reproducibility; the same data were - 7.0, 61.4, 63.4 mm2, and only 3%, respectively, for the inter-observer analysis.

CONCLUSION: Our semi-automatic tool for BAC quantification showed a poor reproducibility. These results pointed out that the human identification of BAC represents the main source of variability. Further research is needed to translate BAC quantification into clinical practice.

来源:Radiol Med,查看原文。版权归原作者所有。