现代前列腺磁共振扩散加权成像序列的图像质量与几何变形比较

Image Quality and Geometric Distortion of Modern Diffusion-Weighted Imaging Sequences in Magnetic Resonance Imaging of the Prostate.
PubMed 文献 · Invest Radiol · 2018-01-01

在10名健康志愿者3T磁共振上比较4种前列腺扩散加权成像序列,发现单次激发自旋回波EPI与iShim-EPI在分辨率和信噪比上倾向优于rs-EPI与sTX-EPI,各序列间几何变形无统计学显著差异。

要点

  • ss-EPI与iShim-EPI分辨率主观评分倾向优于rs-EPI与sTX-EPI。
  • iShim-EPI信噪比显著高于sTX-EPI,其余序列间几何变形无显著差异。
  • sTX-EPI在外周带与移行带的ADC值低于ss-EPI和rs-EPI。

核心结论

  • ss-EPI与iShim-EPI相比rs-EPI和sTX-EPI呈现出图像质量与信噪比更优的趋势,各序列间几何变形无统计学显著差异。

研究设计与样本

  • 在3T磁共振上对10名健康志愿者(平均年龄26.1±3.8岁,体重指数23.2±3.0 kg/m)采集轴位T2加权涡轮自旋回波图像及4种轴位DWI回波平面成像序列:单次激发自旋回波EPI、读出分段多激发EPI、选择性激励缩小视野EPI和原型单次激发层面特异性匀场EPI,采用可比成像参数与相近采集时间。

诊断标准与关键数值

  • 两名放射科医生对DWI以5点Likert量表独立评分分辨率、前列腺包膜分界和带区解剖,读者间一致性组内相关系数0.71-0.79。
  • 分辨率平均Likert评分:ss-EPI 4.25 vs rs-EPI 3.8(P=0.031),ss-EPI 4.25 vs sTX-EPI 3.3(P=0.046)。
  • 分辨率平均Likert评分:iShim-EPI 4.4 vs rs-EPI 3.8(P=0.031),iShim-EPI 4.4 vs sTX-EPI 3.3(P=0.047)。
  • 包膜分界与带区解剖在各序列间无显著差异。
  • 信噪比(均值±标准差):iShim-EPI 28.13±8.21 vs sTX-EPI 14.96±2.4(P=0.015)。

临床应用与阅片要点

  • 在相近采集时间与可比成像参数下,ss-EPI与iShim-EPI的扩散加权图像分辨率和信噪比表现倾向更优。
  • sTX-EPI测得的外周带与移行带ADC值低于ss-EPI和rs-EPI,解读ADC值时需结合所用序列。
  • 各DWI序列在精阜水平的前列腺左右径与前后径几何变形无统计学显著差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The aim of this study was to compare qualitative and quantitative image quality and geometric distortion of 4 magnetic resonance diffusion-weighted imaging (DWI) sequences of the prostate using comparable imaging parameters and similar acquisition times.

METHODS AND MATERIALS: Axial T2-weighted turbo spin echo images and axial DWI echo-planar imaging (EPI) sequences, including single-shot spin-echo (ss-EPI), readout-segmented multishot (rs-EPI), selective excitation-reduced field of view (sTX-EPI), and prototype single-shot technique applying slice-specific shimming (iShim-EPI) sequences, were acquired at 3 T in 10 healthy volunteers (mean age, 26.1 ± 3.8 years; body mass index, 23.2 ± 3.0 kg/m). Two radiologists, blinded to the type of DWI, independently rated DWIs on a 5-point Likert scale regarding subjective image quality features (resolution, demarcation of prostate capsule, zonal anatomy). Interreader agreement was assessed using the intraclass correlation coefficient. Signal-to-noise ratio (SNR) and apparent diffusion coefficient (ADC) values were assessed separately in the peripheral and transitional zone. For the analysis of geometric distortion, the diameter of the prostate from left to right and from anterior to posterior was measured at the level of the verumontanum on b-1000 DWIs and on T2-weighted turbo spin echo images. Differences were compared using the Wilcoxon rank sum test for qualitative parameters, analysis of variance, and Friedman test for quantitative parameters. A P value of less than 0.05 was considered significant, with correction for multiple comparisons.

RESULTS: Interreader agreement was good to excellent (intraclass correlation coefficient, 0.71-0.79) for all qualitative features. Subjective image quality regarding "resolution" was significantly better for ss-EPI than rs-EPI (mean Likert score, 4.25 vs 3.8; P = 0.031) and sTX-EPI (4.25 vs 3.3; P = 0.046) and for iShim-EPI as compared with rs-EPI (4.4 vs 3.8; P = 0.031) and sTX-EPI (4.4 vs 3.3; P = 0.047). There was no significant difference regarding capsule demarcation and zonal anatomy. Signal-to-noise ratio was significantly higher in iShim-EPI than sTX-EPI (SNR ± standard deviation [SD], 28.13 ± 8.21 vs 14.96 ± 2.4; P = 0.015). The ADC values were lower for the peripheral zone in the sTX-EPI than in the ss-EPI (ADC ± SD, 1002.94 ± 83.51 vs 1165.05 ± 115.64; P = 0.013) and the rs-EPI (1002.94 ± 83.51 vs 1244.40 ± 89.95; P = 0.0012) and in the transitional zone in the sTX-EPI compared with the rs-EPI (874.50 ± 200.72 vs 1261.47 ± 179.23; P = 0.0021). There were no statistically significant differences in geometric distortion between all DWI sequences.

CONCLUSIONS: Single-shot spin-echo EPI and iShim-EPI showed a tendency toward superior image quality and SNR compared with rs-EPI and sTX-EPI with no significant differences in geometric distortion.