迭代重建(AIDR 3D)与常规重建在腹盆腔CT中的辐射剂量与图像质量比较
回顾性纳入99例于不同时间点接受腹盆腔CT门静脉期检查、图像分别经滤波反投影与AIDR 3D重建的患者,比较辐射剂量与图像质量。AIDR 3D使CTDIvol、有效剂量、SSDE分别平均降低62.5%、58%、63%(p<0.001),主观图像质量除BMI 20-24.9患者锐利度外均非劣于滤波反投影(p<0.001),客观图像质量结果不一。
要点
- AIDR 3D重建使腹盆腔CT的CTDIvol平均降低62.5%、有效剂量降低58%、SSDE降低63%(p<0.001)。
- 主观图像质量除BMI 20-24.9患者的锐利度外均非劣于滤波反投影(p<0.001)。
- 客观图像质量指标结果不一,需结合主观评分综合判断。
核心结论
- AIDR 3D使腹盆腔CT检查辐射剂量显著降低,总体图像质量未见损失。
研究设计与样本
- 回顾性研究,纳入99例于不同时间点接受腹盆腔CT门静脉期检查、图像分别经滤波反投影与AIDR 3D重建的患者;由两名放射科医师盲法对轴位图像的主观质量(整体质量、锐利度、噪声、可接受性)评分;客观指标包括肝、主动脉、椎旁肌平均衰减的SD(作为图像噪声测量)及肝、主动脉的对比噪声比与信噪比;辐射剂量参数包括CTDIvol、剂量长度乘积、有效剂量和SSDE;按不同BMI分类比较;采用配对t检验及McNemar配对非劣效检验,p<0.05为有统计学意义。
诊断标准与关键数值
- AIDR 3D与滤波反投影相比,CTDIvol平均降低62.5%、有效剂量平均降低58%、SSDE平均降低63%(p<0.001)。
- 主观图像质量参数在AIDR 3D与滤波反投影间为非劣效(p<0.001),但BMI 20-24.9患者的图像锐利度除外。
- 客观图像质量评估结果不一。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The purpose of this study is to compare radiation dose and image quality of abdominopelvic CT studies reconstructed with iterative and conventional techniques.
MATERIALS AND METHODS: This retrospective study enrolled 99 patients who underwent abdominopelvic CT examinations with the portal venous phase images reconstructed with both filtered back projection and Adaptive Iterative Dose Reduction 3D (AIDR 3D) at different time points. Subjective assessment of image quality was performed by two radiologists who scored axial images for overall quality, sharpness, noise, and acceptability in a blinded fashion. The SD of the mean attenuation of the liver, aorta, and paraspinal muscle (as a measurement of image noise) and contrast-to-noise and signal-to-noise ratios for liver and aorta were used as objective parameters of image quality. Radiation dose parameters included CT dose index volume (CTDIvol), dose-length product, effective dose (ED), and size-specific dose estimate (SSDE). Results were compared for different body mass index (BMI; weight in kilograms divided by the square of height in meters) categories. Paired t test and McNemar paired tests for noninferiority were used, with p < 0.05 considered statistically significant.
RESULTS: We obtained a 62.5% mean reduction in CTDIvol, a 58% mean reduction in ED, and a 63% mean reduction in SSDE when AIDR 3D was used (p < 0.001). Subjective parameters of image quality were considered noninferior for AIDR 3D studies compared with filtered back projection (p < 0.001), except for the sharpness of images of patients with BMI 20-24.9. Variable results were found regarding objective assessment of image quality.
CONCLUSION: AIDR 3D allowed a significant reduction in radiation dose of abdominopelvic CT examinations without a loss of image quality in general.