3特斯拉磁共振成像无创测量肝铁浓度:与活检对照验证
前瞻性纳入105例接受肝活检并测定生化肝铁浓度(LICb)的患者,均行3T多回波梯度回波序列扫描。LICb与R2*(r=0.95)及信号强度比(SIR)法(r=0.92)均强相关;R2*和SIR检出铁过载的敏感度/特异度分别为0.96/0.93和0.92/0.95。结论:低铁过载时R2*定量精确,高铁过载时推荐SIR法。
要点
- 3T MRI的R2*与SIR法均可无创定量肝铁浓度,与活检生化值强相关。
- R2*法在LICb低于130 μmol/g时差异最小,高铁过载时SIR法差异最小。
- R2*和SIR检出肝铁过载的敏感度/特异度分别为0.96/0.93和0.92/0.95。
核心结论
- 3T MRI的R2*法对较低铁过载提供精确的肝铁浓度定量,但高铁过载时推荐SIR法以克服R2*的局限性;作者软件联合使用两种方法并选择最佳者。
研究设计与样本
- 前瞻性纳入105例接受肝活检并测定生化肝铁浓度(LICb)的患者,均行3T MRI屏气多回波梯度回波序列(mGRE)扫描,以活检生化LICb为参考标准。
诊断标准与关键数值
- LICb与R2*相关系数r=0.95(p<0.001),与LICSIR相关系数r=0.92(p<0.001);与LICb比较,LICR2*检出肝铁过载的敏感度/特异度为0.96/0.93,LICSIR为0.92/0.95。
临床应用与阅片要点
- 低铁过载时R2*法定量精确,高铁过载时推荐SIR法;作者软件联合使用两种方法并选择最佳者。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate the performance and limitations of the R2* and signal intensity ratio (SIR) methods for quantifying liver iron concentration (LIC) at 3 T.
METHODS: A total of 105 patients who underwent a liver biopsy with biochemical LIC (LICb) were included prospectively. All patients underwent a 3-T MRI scan with a breath-hold multiple-echo gradient-echo sequence (mGRE). LIC calculated by 3-T SIR algorithm (LICSIR) and by R2* (LICR2*) were correlated with LICb. Sensitivity and specificity were calculated. The comparison of methods was analysed for successive classes.
RESULTS: LICb was strongly correlated with R2* (r = 0.95, p < 0.001) and LICSIR (r = 0.92, p < 0.001). In comparison to LICb, LICR2* and LICSIR detect liver iron overload with a sensitivity/specificity of 0.96/0.93 and 0.92/0.95, respectively, and a bias ± SD of 7.6 ± 73.4 and 14.8 ± 37.6 μmol/g, respectively. LICR2* presented the lowest differences for patients with LICb values under 130 μmol/g. Above this value, LICSIR has the lowest differences.
CONCLUSIONS: At 3 T, R2* provides precise LIC quantification for lower overload but the SIR method is recommended to overcome R2* limitations in higher overload. Our software, available at www.mrquantif.org , uses both methods jointly and selects the best one.
KEY POINTS: • Liver iron can be accurately quantified by MRI at 3 T • At 3 T, R2* provides precise quantification of slight liver iron overload • At 3 T, SIR method is recommended in case of high iron overload • Slight liver iron overload present in metabolic syndrome can be depicted • Treatment can be monitored with great confidence.
引用
- d'Assignies G, Paisant A, Bardou-Jacquet E, et al. Non-invasive measurement of liver iron concentration using 3-Tesla magnetic resonance imaging: validation against biopsy. Eur Radiol. 2017. PMID: 29178028. doi:10.1007/s00330-017-5106-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。