双能CT技术中基于原始数据的有效原子序数和单色CT值对对比剂的准确性
研究使用组织表征体模和不同浓度碘对比剂,评估双能CT基于原始数据获得的有效原子序数(Zeff)值和单色CT值的准确性。结果显示组织等效体模中Zeff和单色CT值相当准确,对比剂Zeff值在临床动态CT碘浓度范围内与理论值吻合良好,人体组织和碘对比剂的最佳单色能量为70 keV。
要点
- 基于原始数据的Zeff值在组织等效体模中的准确性高于基于图像的数据。
- 对比剂Zeff值与理论值在临床动态CT碘浓度范围内吻合良好,平均差异在11.2%以内。
- 人体组织和碘对比剂的最佳单色能量为70 keV。
核心结论
- 基于原始数据的双能CT所生成的对比剂Zeff值和单色CT值在临床条件下可能足够准确,且人体组织和碘对比剂的最佳单色能量为70 keV。
研究设计与样本
- 研究使用组织表征体模和不同浓度的碘对比剂,比较Zeff和单色CT值的测量值与理论值或制造商提供值。
诊断标准与关键数值
- 组织表征体模中肺(吸气)插入物的Zeff平均差异为81.3%,其余Zeff平均差异在8.4%以内。
- 不同浓度碘对比剂的Zeff平均差异在11.2%以内。
- 不同浓度碘对比剂的单色CT测量值与理论值之间的差异随单色能量降低而增大。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the accuracy of raw-data-based effective atomic number (Zeff) values and monochromatic CT numbers for contrast material of varying iodine concentrations, obtained using dual-energy CT.
METHODS: We used a tissue characterization phantom and varying concentrations of iodinated contrast medium. A comparison between the theoretical values of Zeff and that provided by the manufacturer was performed. The measured and theoretical monochromatic CT numbers at 40-130 keV were compared.
RESULTS: The average difference between the Zeff values of lung (inhale) inserts in the tissue characterization phantom was 81.3% and the average Zeff difference was within 8.4%. The average difference between the Zeff values of the varying concentrations of iodinated contrast medium was within 11.2%. For the varying concentrations of iodinated contrast medium, the differences between the measured and theoretical monochromatic CT values increased with decreasing monochromatic energy. The Zeff and monochromatic CT numbers in the tissue characterization phantom were reasonably accurate.
CONCLUSION: The accuracy of the raw-data-based Zeff values was higher than that of image-based Zeff values in the tissue-equivalent phantom. The accuracy of Zeff values in the contrast medium was in good agreement within the maximum SD found in the iodine concentration range of clinical dynamic CT imaging. Moreover, the optimum monochromatic energy for human tissue and iodinated contrast medium was found to be 70 keV. Advances in knowledge: The accuracy of the Zeff values and monochromatic CT numbers of the contrast medium created by raw-data-based, dual-energy CT could be sufficient in clinical conditions.