320排CT虚拟单色图像质量:体模研究评估迭代重建与体型的影响

Image quality of virtual monochromatic images obtained using 320-detector row CT: A phantom study evaluating the effects of iterative reconstruction and body size.
PubMed 文献 · Eur J Radiol · 2017-08-19

该体模研究比较了320排CT虚拟单色光谱图像与120-kVp多色图像的图像质量,模拟三种体型并采用FBP和AIDR-3D重建。结果显示FBP重建时VMS图像质量多低于或类似120-kVp图像,而AIDR-3D重建时约70-keV的VMS图像在多数设置下信噪比和对比噪声比高于120-kVp图像。

要点

  • 行320排CT双能量扫描时,采用AIDR-3D重建的约70-keV虚拟单色图像在多数剂量和体型条件下信噪比和对比噪声比高于120-kVp图像。
  • 采用FBP重建时,虚拟单色图像的信噪比和对比噪声比在多数剂量设置下低于或类似120-kVp图像。
  • 体模研究提示重建算法(AIDR-3D vs FBP)对虚拟单色图像质量的影响大于体型和剂量设置的差异。

核心结论

  • 使用FBP重建时,VMS图像质量倾向于低于120-kVp图像;使用AIDR-3D重建时,约70-keV的VMS图像质量高于120-kVp图像。

研究设计与样本

  • 体模研究,使用模拟三种患者体型的躯干体模,内含四支注射器分别填充水或不同浓度对比剂(5、10、15 mgI/mL),行双能量(80/135-kVp)和单能量(120-kVp)扫描,剂量设置为20 mGy、12 mGy和6 mGy,VMS图像以1-keV间隔生成(范围35-135 keV),分别用FBP和AIDR-3D重建。

诊断标准与关键数值

  • 评估指标为信噪比(SNR)和对比噪声比(CNR);FBP重建下多数剂量设置VMS图像的SNR和CNR低于或类似120-kVp图像;AIDR-3D重建下多数设置70-keV VMS图像的SNR和CNR高于120-kVp图像。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the image quality between virtual monochromatic spectral (VMS) images obtained using 320-row detector CT and polychromatic 120-kVp images reconstructed with or without iterative reconstruction using various phantom sizes.

MATERIALS AND METHODS: Torso phantoms simulating three patient sizes and containing four syringes filled with water or different contrast media (5, 10, 15mgI/mL15mgI/ml) were used. The phantoms were scanned using dual-energy (80/135-kVp) and single-energy (120-kVp) protocols at different settings (20mGy, 12mGy, and 6mGy). VMS images were generated at 1-keV intervals (range, 35-135keV). Both the VMS images and the single-energy 120-kVp images were reconstructed using filtered back projection (FBP) and adaptive iterative dose reduction 3D (AIDR-3D). The signal-to-noise ratio (SNR), and the contrast-to-noise ratio (CNR) were assessed.

RESULTS: Using FBP reconstruction, the SNR and CNR of the VMS images were lower than or similar to those of the 120-kVp images for most dose settings. Using AIDR-3D reconstruction, however, the 70-keV VMS images had higher SNRs and CNRs than the 120-kVp images at most settings.

CONCLUSIONS: The image quality of VMS images with FBP reconstruction tended to be lower than that of the 120-kVp images. With the use of AIDR-3D, however, approximately 70-keV VMS images had a higher image quality than the 120-kVp images.

来源:Eur J Radiol,查看原文。版权归原作者所有。