文献速递骨肌儿科影像介入MR

磁共振成像环境下机器人辅助长骨活检:工作流程与临床前研究

Robotically Assisted Long Bone Biopsy Under MRI Imaging: Workflow and Preclinical Study.
PubMed 文献 · Acad Radiol · 2017-10-23

研究团队开发了一种磁共振成像(MRI)兼容机器人,用于在MRI环境下对长骨进行活检,目标是在诊断儿童骨肿瘤和骨感染时减少创伤并消除辐射暴露。在西门子Aera 1.5T扫描仪上使用长骨体模进行精度研究,共8次瞄准尝试,平均三维瞄准误差为1.39 mm(标准差0.40 mm),所有目标均成功到达,平均每次瞄准尝试耗时32分钟(含机器人安装时间)。

要点

  • MRI兼容机器人具有3个自由度,采用远程运动中心机构定向针导引器,在1.5T MRI环境中辅助精准瞄准骨体模。
  • 体模精度研究中平均三维瞄准误差为1.39 mm(标准差0.40 mm),所有目标均成功到达。
  • 工作流程分析显示平均每次瞄准尝试耗时32分钟,包括机器人安装时间。

核心结论

  • 结果表明,该机器人系统能够在MRI环境下辅助放射科医生精确瞄准骨体模。

研究设计与样本

  • 临床前体模精度研究,使用MRI兼容机器人(MR安全、3个自由度、远程运动中心机构)在西门子Aera 1.5T扫描仪上对长骨体模进行瞄准,体模上钻有4个靶孔,每个靶孔以略微倾斜的角度接近两次,共8次尝试。

诊断标准与关键数值

  • 平均三维瞄准误差为1.39 mm(标准差0.40 mm),所有目标均成功到达。

临床应用与阅片要点

  • 作者认为该机器人系统在临床应用中具有若干潜在优势,包括能够在MRI等中心工作以及作为稳定且精确的导引。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Our research team has developed a magnetic resonance imaging (MRI)-compatible robot for long bone biopsy. The robot is intended to enable a new workflow for bone biopsy in pediatrics under MRI imaging. Our long-term objectives are to minimize trauma and eliminate radiation exposure when diagnosing children with bone cancers and bone infections. This article presents our robotic systems, phantom accuracy studies, and workflow analysis.

MATERIALS AND METHODS: This section describes several aspects of our work including the envisioned clinical workflow, the MRI-compatible robot, and the experimental setup. The workflow consists of five steps and is intended to enable the entire procedure to be completed in the MRI suite. The MRI-compatible robot is MR Safe, has 3 degrees of freedom, and a remote center of motion mechanism for orienting a needle guide. The accuracy study was done in a Siemens Aera 1.5T scanner with a long bone phantom. Four targeting holes were drilled in the phantom.

RESULTS: Each target was approached twice at slightly oblique angles using the robot needle guide for a total of eight attempts. A workflow analysis showed the average time for each targeting attempt was 32 minutes, including robot setup time. The average 3D targeting error was 1.39 mm with a standard deviation of 0.40 mm. All of the targets were successfully reached.

CONCLUSION: The results showed the ability of the robotic system in assisting the radiologist to precisely target a bone phantom in the MRI environment. The robot system has several potential advantages for clinical application, including the ability to work at the MRI isocenter and serve as a steady and precise guide.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Cleary K, Lim S, Jun C, et al. Robotically Assisted Long Bone Biopsy Under MRI Imaging: Workflow and Preclinical Study. Acad Radiol. 2017. PMID: 29074334. doi:10.1016/j.acra.2017.08.008

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。