儿科机构中低剂量C臂计算机断层摄影的应用
该研究回顾性分析5年内222例采用机构特定低剂量方案的C臂CT引导操作,常见为颞下颌关节和骶髂关节注射及硬化治疗;C臂CT占操作总剂量面积乘积的9.9%至80.7%,结论认为低剂量技术可在保持介入图像质量的同时减少儿童辐射暴露,可作为辅助影像工具。
要点
- 低剂量C臂CT采用5或8秒Dyna旋转、248/396幅投影图像及0.1-0.17 μGy/投影的探测器剂量,配合0.3/0.6/0.9 mm铜滤过。
- 常见操作包括颞下颌关节和骶髂关节注射(48.6%)及硬化治疗(34.2%),也用于困难经皮穿刺如盲肠造口、胃造口、异物取出和胸腔穿刺。
- C臂CT占操作总剂量面积乘积的9.9%至80.7%,可作为常规透视或超声引导困难时的辅助影像工具。
核心结论
- 低剂量技术可在减少儿童辐射暴露的同时保持介入图像质量,该技术在儿科介入放射学中有多种应用,可考虑作为多种操作中的辅助影像工具,尤其当常规透视或超声引导下经皮穿刺困难时。
研究设计与样本
- 5年回顾性研究,纳入222例采用机构特定低剂量方案的C臂CT引导操作,方案为5或8秒Dyna旋转、248/396幅投影图像/采集、探测器剂量0.1-0.17 μGy/投影、铜滤过0.3/0.6/0.9 mm。
诊断标准与关键数值
- 低剂量方案参数:5或8秒Dyna旋转,248/396幅投影图像/采集,探测器剂量0.1-0.17 μGy/投影,铜滤过0.3/0.6/0.9 mm。
- C臂CT占操作总剂量面积乘积的范围为9.9%至80.7%。
- 最常见操作:颞下颌关节和骶髂关节注射占48.6%,硬化治疗占34.2%。
临床应用与阅片要点
- 作者认为该技术可作为辅助影像工具用于多种操作,尤其当常规透视或超声引导下经皮穿刺困难时。
- C臂CT可用于盲肠造口、胃造口置入、异物取出和胸腔穿刺等不常见应用中的困难经皮穿刺。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Reduced-dose C-arm computed tomography (CT) uses flat-panel detectors to acquire real-time 3-D images in the interventional radiology suite to assist with anatomical localization and procedure planning.
OBJECTIVE: To describe dose-reduction techniques for C-arm CT at a pediatric institution and to provide guidance for implementation.
MATERIALS AND METHODS: We conducted a 5-year retrospective study on procedures using an institution-specific reduced-dose protocol: 5 or 8 s Dyna Rotation, 248/396 projection images/acquisition and 0.1-0.17 μGy/projection dose at the detector with 0.3/0.6/0.9-mm copper (Cu) filtration. We categorized cases by procedure type and average patient age and calculated C-arm CT and total dose area product (DAP).
RESULTS: Two hundred twenty-two C-arm CT-guided procedures were performed with a dose-reduction protocol. The most common procedures were temporomandibular and sacroiliac joint injections (48.6%) and sclerotherapy (34.2%). C-arm CT was utilized in cases of difficult percutaneous access in less common applications such as cecostomy and gastrostomy placement, foreign body retrieval and thoracentesis. C-arm CT accounted for between 9.9% and 80.7% of the total procedural DAP.
CONCLUSION: Dose-reducing techniques can preserve image quality for intervention while reducing radiation exposure to the child. This technology has multiple applications within pediatric interventional radiology and can be considered as an adjunctive imaging tool in a variety of procedures, particularly when percutaneous access is challenging despite routine fluoroscopic or ultrasound guidance.
引用
- Acord M, Shellikeri S, Vatsky S, et al. Reduced-dose C-arm computed tomography applications at a pediatric institution. Pediatr Radiol. 2017. PMID: 28983677. doi:10.1007/s00247-017-3964-0
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。