本研究扩展了高帧率超声成像测速技术,采用非相干系综相关方法实现更稳健的流速定量映射,并生成时空壁面剪切应力分布。仿真显示该方法流速估计误差较相干相关方法降低最多10倍,稳态流速与壁面剪切率平均误差分别为3.2%和8.6%,搏动条件下分别为3.0%和10.6%。在正常与病变颈动脉分叉体模的体外研究中展示了血流模式与壁面剪切率分布的相对差异。

要点

  • 非相干系综相关流速估计误差较相干相关方法降低最多10倍。
  • 壁面跟踪偏差200 µm可致壁面剪切率误差高达60%,高帧率对比增强成像可减轻此影响。
  • 稳态下流速与壁面剪切率平均误差分别为3.2%和8.6%。

核心结论

  • 高帧率超声成像测速技术可作为一种无创工具定量测量时空流速与壁面剪切应力分布。

研究设计与样本

  • 仿真模型将超声模拟器与解析流场解耦合以评估准确性,并在正常与病变两个颈动脉分叉体模上进行体外研究。

关键影像表现/征象

  • 正常与病变颈动脉分叉体模在血流模式和壁面剪切率分布上表现出相对差异。

诊断标准与关键数值

  • 稳态条件下流速测量平均误差3.2%,壁面剪切率测量平均误差8.6%。
  • 搏动条件下流速测量平均误差3.0%,壁面剪切率测量平均误差10.6%。
  • 壁面跟踪偏差200 µm可致壁面剪切率误差高达60%。
  • 高帧率成像帧率可达10 kHz,使用对比剂可改善壁-流体边界区分。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

In this study, a technique for high-frame-rate ultrasound imaging velocimetry (UIV) is extended first to provide more robust quantitative flow velocity mapping using ensemble correlation of images without coherent compounding, and second to generate spatio-temporal wall shear stress (WSS) distribution. A simulation model, which couples the ultrasound simulator with analytical flow solution, was implemented to evaluate its accuracy. It is shown that the proposed approach can reduce errors in velocity estimation by up to 10-fold in comparison with the coherent correlation approach. Mean errors (ME) of 3.2% and 8.6% were estimated under a steady flow condition, while 3.0% and 10.6% were found under a pulsatile condition for the velocity and wall shear rate (WSR) measurement, respectively. Appropriate filter parameters were selected to constrain the velocity profiles before WSR estimations and the effects of incorrect wall tracking were quantified under a controlled environment. Although accurate wall tracking is found to be critical in WSR measurement (as a 200 µm deviation from the wall may yield up to a 60% error), this can be mitigated by HFR imaging (of up to 10 kHz) with contrast agents, which allow for improved differentiation of the wall-fluid boundaries. In vitro investigations on two carotid bifurcation phantoms, normal and diseased, were conducted, and their relative differences in terms of the flow patterns and WSR distribution were demonstrated. It is shown that high-frame-rate UIV technique can be a non-invasive tool to measure quantitatively the spatio-temporal velocity and WSS distribution.