透视引导下注射的剂量降低:体模模拟与患者操作
研究通过体模模拟和患者注射评估剂量最小化技术(DMT)对透视引导下关节、滑囊和腱鞘注射的辐射剂量影响。体模模拟显示DMT使总DAP从191.7降至18.7 μGy·m2,入口剂量从10.2降至3.6 mGy。1,094例患者注射中,DMT后透视时间、剂量和DAP均显著降低,且在肩关节和髋关节注射亚组中仍具统计学意义(P < 0.05)。
要点
- 使用简单易学的DMT可降低透视引导下注射的透视时间、有效剂量、入口剂量和DAP。
- 1,094例注射中,DMT后透视时间从0.7 min降至0.2 min,剂量从5.6 mGy降至1.9 mGy,DAP从56.9 μGy·m2降至19.1 μGy·m2。
- 在肩关节和髋关节注射亚组中,DMT的剂量降低仍具统计学意义(P < 0.05)。
核心结论
- 使用简单易学的技术可降低透视时间、有效剂量、入口剂量和DAP,从而使患者和放射科医生受益。
研究设计与样本
- 研究经IRB批准并符合HIPAA,包括体模模拟和1,094例由三名操作者在同一透视室进行的关节、滑囊和腱鞘注射的DMT前后评估,使用非配对t检验,P > 0.05视为有统计学意义。
诊断标准与关键数值
- 体模模拟中,无DMT与有DMT相比,总DAP为191.7 vs 18.7 μGy·m2,入口剂量为10.2 vs 3.6 mGy。
- 所有注射中,DMT后透视时间从0.7 min降至0.2 min,剂量从5.6 mGy降至1.9 mGy,DAP从56.9 μGy·m2降至19.1 μGy·m2。
- 按两种最常见注射(肩关节和髋关节注射)分层后,DMT的降低仍具统计学意义(P < 0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To demonstrate fluoroscopy dose reduction through both simulated injections on a phantom and patient injections.
MATERIALS AND METHODS: Our study was IRB-approved and HIPAA-compliant. Simulation on a phantom was used to estimate effective dose, entrance dose, and organ doses for hip joint injections without and with dose minimization technique (DMT). Additionally, 1,094 joint, bursae, and tendon sheath injections performed by three operators in the same fluoroscopy suite were evaluated both before and after application of DMT. Fluoroscopy time (FT), dose, and dose area product (DAP) of injections were compared using unpaired t-tests with P > 0.05 considered statistically significant.
RESULTS: For the phantom simulation comparing injections without DMT and with DMT, the total DAP was 191.7 vs 18.7 μGy·m2, and the entrance dose was 10.2 vs 3.6 mGy, respectively. For both men and women, DMT reduces effective dose and organ doses. For all injections, the FT (0.7 to 0.2 min), dose (5.6 to 1.9 mGy), and DAP (56.9 to 19.1 μGy·m2) for all three operators decreased with DMT and remained statistically significant when stratified by the two most common injections, glenohumeral and hip joint injections (P < 0.05).
CONCLUSIONS: FT, effective dose, entrance dose, and DAP can be reduced with the use of simple easy-to-learn techniques, which will benefit both the patient and the radiologist.
引用
- Chang CY, Simeone FJ, DeLorenzo MC, et al. Dose reduction for fluoroscopically guided injections: phantom simulation and patient procedures. Skeletal Radiol. 2017. PMID: 29027573. doi:10.1007/s00256-017-2788-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。