声辐射力脉冲成像定量评估儿童桥本甲状腺炎患者甲状腺弹性的研究

Acoustic Radiation Force Impulse Quantification in the Evaluation of Thyroid Elasticity in Pediatric Patients With Hashimoto Thyroiditis.
PubMed 文献 · J Ultrasound Med · 2017-10-24

研究纳入26例桥本甲状腺炎患儿与26名健康志愿者,采用声辐射力脉冲成像测量双侧甲状腺剪切波速度。桥本组剪切波速度均值1.67±0.63 m/s显著高于对照组1.30±0.13 m/s(P<.001)。以1.41 m/s为最佳截断值,敏感度73.1%、特异度80.8%、AUC 0.806;桥本组剪切波速度与抗甲状腺过氧化物酶呈正相关(Pearson r=0.46,P=.038)。

要点

  • 儿童桥本甲状腺炎甲状腺剪切波速度均值1.67±0.63 m/s,显著高于健康对照组1.30±0.13 m/s。
  • 最佳截断值1.41 m/s,敏感度73.1%、特异度80.8%、AUC 0.806(P<.001)。
  • 桥本组剪切波速度与抗甲状腺过氧化物酶正相关(Pearson r=0.46,P=.038)。

核心结论

  • 声辐射力脉冲弹性成像显示桥本甲状腺炎患儿与健康儿童甲状腺硬度存在显著差异,在标准超声评估后立即行该检查可能预测慢性自身免疫性甲状腺炎。

研究设计与样本

  • 研究纳入26例桥本甲状腺炎患者和26名6至17岁健康志愿者,评估两组双侧甲状腺叶的剪切波速度值。

关键影像表现/征象

  • 桥本甲状腺炎组甲状腺剪切波速度均值±标准差为1.67±0.63 m/s,显著高于对照组1.30±0.13 m/s(P<.001)。
  • 两组间甲状腺左、右叶剪切波速度无显著差异。
  • 两组间腺体大小、腺体血供及左甲状腺素治疗无显著差异。

诊断标准与关键数值

  • 受试者工作特征曲线分析显示最佳截断值为1.41 m/s,敏感度73.1%、特异度80.8%、阳性预测值79.2%、阴性预测值75.0%,曲线下面积0.806(P<.001)。

临床应用与阅片要点

  • 在标准超声评估后立即行声辐射力脉冲弹性成像测量甲状腺剪切波速度,可能有助于预测儿童慢性自身免疫性甲状腺炎。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the parenchymal elasticity of the thyroid gland with acoustic radiation force impulse imaging in pediatric patients with Hashimoto thyroiditis and to compare it with healthy volunteers.

METHODS: Twenty-six patients with Hashimoto thyroiditis and 26 healthy volunteers between 6 and 17 years were included. The shear wave velocity (SWV) values of both thyroid lobes in both groups were evaluated.

RESULTS: The age and sex characteristics of the controls and patients with Hashimoto thyroiditis were similar. The SWV of the thyroid gland in patients with Hashimoto thyroiditis (mean ± SD, 1.67 ± 0.63 m/s) was significantly higher than that in the control group (1.30 ± 0.13 m/s; P < .001). There was no significant difference between the thyroid lobes in both groups. A receiver operating characteristic curve analyses showed an optimal cutoff value of 1.41 m/s, with 73.1% sensitivity, 80.8% specificity, a 79.2 % positive predictive value, and a 75.0% negative predictive value (area under the curve, 0.806; P < .001). In patients with Hashimoto thyroiditis, there was a positive correlation between the SWV values versus anti-thyroperoxidase (Pearson r = 0.46; P = .038). There were no correlations between age, body mass index, thyroid function test results, and anti-thyroglobulin values and versus SWV values. Also, no significant differences were seen between the groups for gland size, gland vascularity, and l-thyroxine treatment.

CONCLUSIONS: Acoustic radiation force impulse elastography showed a significant difference in the stiffness of the thyroid gland between children with Hashimoto thyroiditis and the healthy group. Using acoustic radiation force impulse elastography immediately after a standard ultrasound evaluation may predict chronic autoimmune thyroiditis.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Yucel S, Ceyhan Bilgici M, Kara C, et al. Acoustic Radiation Force Impulse Quantification in the Evaluation of Thyroid Elasticity in Pediatric Patients With Hashimoto Thyroiditis. J Ultrasound Med. 2017. PMID: 29064111. doi:10.1002/jum.14459

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。