文献速递乳腺乳腺X线

对比增强数字乳腺摄影上模拟原位癌的皮肤污染伪影

Contamination artifact that mimics in-situ carcinoma on contrast-enhanced digital mammography.
PubMed 文献 · Eur J Radiol · 2017-08-05

本文报道5例对比增强数字乳腺摄影(CEDM)图像上出现疑似钙化及非肿块强化、酷似原位癌但后续无病变证据的病例,回顾性分析未处理及处理后图像与成像参数,并通过双能乳腺模体实验证实皮肤表面碘对比剂污染可复制该伪影。

要点

  • CEDM增强图像上疑似钙化及非肿块强化可为皮肤碘对比剂污染所致,酷似原位癌。
  • 该伪影在接近碘K边缘时信号衰减,可借此与真性病变鉴别。
  • 模体实验证实皮肤表面碘对比剂污染可复制类似原位癌的明显强化表现。

核心结论

  • 皮肤表面碘对比剂污染可在增强后数字乳腺摄影上产生伪影,其表现酷似原位癌。

研究设计与样本

  • 病例系列纳入5例CEDM图像上疑似原位癌但后续无病变证据的患者,回顾性分析未处理图像数据、处理后图像及成像参数,并在两个乳腺模体上以不同程度上皮肤碘对比剂污染行双能乳腺摄影技术。

关键影像表现/征象

  • CEDM图像上表现为疑似钙化及非肿块强化,酷似原位癌。
  • 时间分析证实该可疑发现既非压迫板异常也非探测器异常。
  • 低能与高能图像比较显示可疑发现在接近碘K边缘时衰减,提示对比剂污染。
  • 乳腺模体表面涂抹碘对比剂可复制该伪影,呈类似原位癌的明显强化表现。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Contrast-enhanced digital mammography (CEDM) is an advanced breast imaging technique using iodinated intravenous contrast to detect breast cancer. This article describes imaging features of a skin contamination artifact on CEDM that mimics in-situ carcinoma in a case series.

MATERIALS AND METHODS: Five patients were identified whose CEDM images demonstrated apparent calcifications and non-mass enhancement suspicious for in-situ carcinoma, with no subsequent evidence of disease. Retrospective image analysis was performed on the unprocessed image data, processed images, and imaging parameters. Dual-energy mammographic technique was performed on two breast phantoms with varying degrees of topical contrast contamination.

RESULTS: Temporal analysis confirmed the suspicious finding was neither an abnormality of the compression paddle nor the receptor. Comparison of LE and HE images demonstrated the suspicious finding attenuated near the K-edge of iodine, suggesting contrast contamination. Iodinated contrast applied to the surface of breast phantoms replicated the artifact, with a pattern of apparent enhancement similar to the appearance of in-situ carcinoma.

CONCLUSION: Skin contamination with iodinated contrast can result in an artifact on post-contrast digital mammography that mimics the appearance of in-situ carcinoma.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Gluskin J, Click M, Fleischman R, et al. Contamination artifact that mimics in-situ carcinoma on contrast-enhanced digital mammography. Eur J Radiol. 2017. PMID: 28987661. doi:10.1016/j.ejrad.2017.08.002

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。