儿童及年轻淋巴瘤患者通过联合正电子发射断层扫描/计算机断层扫描(PET/CT)与诊断性CT降低辐射剂量

Radiation dose reduction through combining positron emission tomography/computed tomography (PET/CT) and diagnostic CT in children and young adults with lymphoma.
PubMed 文献 · Pediatr Radiol · 2017-10-28

该回顾性研究比较了联合PET/诊断性CT与分别采集PET/CT和诊断性CT两种方案在儿童及年轻淋巴瘤患者中的总辐射剂量。联合方案CT部分平均有效剂量6.20±2.69 mSv,分别采集方案为8.17±2.61 mSv(P<0.0001),平均节省24.8±17.8%的总CT有效剂量。图像噪声在两种方案间无统计学显著差异(12.2±1.8 HU vs. 11.7±1.5 HU)。

要点

  • 联合PET/诊断性CT在儿童及年轻淋巴瘤患者中可降低CT有效剂量,平均节省24.8±17.8%。
  • 联合方案CT部分平均有效剂量6.20±2.69 mSv,低于分别采集的8.17±2.61 mSv(P<0.0001)。
  • 两种方案图像噪声无统计学显著差异(12.2±1.8 HU vs. 11.7±1.5 HU)。

核心结论

  • 联合PET/诊断性CT方案在图像质量相似的情况下,为有PET和诊断性CT适应证的儿童及年轻淋巴瘤患者节省辐射剂量。

研究设计与样本

  • 回顾性研究,比较联合PET/诊断性CT与分别采集PET/CT和诊断性CT的总辐射剂量;同一患者可获取时计算两种方案的有效剂量,否则使用体型相似患者的数据;采用t检验比较剂量,双侧单侧检验评估图像质量等效性。

诊断标准与关键数值

  • 联合方案CT部分平均总有效剂量6.20±2.69 mSv,分别采集方案为8.17±2.61 mSv(P<0.0001)。
  • 联合方案平均节省总CT有效剂量24.8±17.8%(2.60±2.51 mSv,范围0.32-4.72 mSv)。
  • 图像噪声在联合与分别采集诊断性CT图像间无统计学显著差异(12.2±1.8 HU vs. 11.7±1.5 HU)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Both [F-18]2-fluoro-2-deoxyglucose positron emission tomography/computed tomography (18F-FDG PET/CT) and diagnostic CT are at times required for lymphoma staging. This means some body segments are exposed twice to X-rays for generation of CT data (diagnostic CT + localization CT).

OBJECTIVE: To describe a combined PET/diagnostic CT approach that modulates CT tube current along the z-axis, providing diagnostic CT of some body segments and localization CT of the remaining body segments, thereby reducing patient radiation dose.

MATERIALS AND METHODS: We retrospectively compared total patient radiation dose between combined PET/diagnostic CT and separately acquired PET/CT and diagnostic CT exams. When available, we calculated effective doses for both approaches in the same patient; otherwise, we used data from patients of similar size. To confirm image quality, we compared image noise (Hounsfield unit [HU] standard deviation) as measured in the liver on both combined and separately acquired diagnostic CT images. We used t-tests for dose comparisons and two one-sided tests for image-quality equivalence testing.

RESULTS: Mean total effective dose for the CT component of the combined and separately acquired diagnostic CT exams were 6.20±2.69 and 8.17±2.61 mSv, respectively (P<0.0001). Average dose savings with the combined approach was 24.8±17.8% (2.60±2.51 mSv [range: 0.32-4.72 mSv]) of total CT effective dose. Image noise was not statistically significantly different between approaches (12.2±1.8 HU vs. 11.7±1.5 HU for the combined and separately acquired diagnostic CT images, respectively).

CONCLUSION: A combined PET/diagnostic CT approach as described offers dose savings at similar image quality for children and young adults with lymphoma who have indications for both PET and diagnostic CT examinations.

来源:Pediatr Radiol,查看原文。版权归原作者所有。