儿科及青年人群中手工弛豫法与半自动扫描仪多回波Dixon技术测量肝脏T2*的一致性
本研究在2015年2月至2016年6月间回顾性分析109例于1.5T行肝脏铁含量评估的临床MRI检查,比较半自动3D多回波Dixon序列与基于梯度回波的多回波弛豫法手工拟合曲线所测肝脏T2*值。两种序列平均T2*值无统计学显著差异(11.7±11.0 ms vs 11.7±10.9 ms,P=0.93),组内相关系数ICC=0.98(CI 0.97-0.98),平均偏差0.0 ms(-4.2至+4.2 ms),提示3D mDixon测量肝脏T2*准确,可能替代2D GRE弛豫法。
要点
- 3D mDixon与GRE法测得肝脏T2*均值无统计学显著差异(11.7±11.0 ms vs 11.7±10.9 ms,P=0.93)。
- 两序列绝对一致性极佳,ICC=0.98(CI 0.97-0.98),平均偏差0.0 ms(-4.2至+4.2 ms)。
- 3D mDixon可测量肝脏T2*且可能替代基于2D GRE的弛豫法,简化临床工作流程。
核心结论
- 3D mDixon测量肝脏T2*准确,可能替代基于2D GRE的弛豫法。
研究设计与样本
- 回顾性分析2015年2月至2016年6月间于1.5T行肝脏铁含量评估的临床MRI检查,共纳入109例患者(site 1=82例,site 2=27例),比较3D mDixon与多回波GRE序列,于四层肝脏中部轴位层面绘制感兴趣区获取平均T2*值。
诊断标准与关键数值
- 两种序列平均T2*值无统计学显著差异(合并均值:GRE 11.7±11.0 ms vs mDixon 11.7±10.9 ms,P=0.93);绝对一致性ICC=0.98(CI 0.97-0.98),平均偏差0.0 ms(-4.2至+4.2 ms)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Commercially available 3D multi-echo Dixon (mDixon) sequences provide parametric maps of liver T2*, obviating manual curve fitting that is often required with conventional gradient recalled echo (GRE)-based multi-echo relaxometry, potentially simplifying clinical work flow.
OBJECTIVE: The purpose of our study was to compare T2* values generated by a 3D mDixon sequence to values generated by GRE-based T2* relaxometry with manual curve fitting in a pediatric and young adult population.
MATERIALS AND METHODS: We reviewed clinical MRI exams performed at 1.5T for liver iron content estimation between February 2015 and June 2016 that included both mDixon and multi-echo GRE pulse sequences. We obtained mean T2* measurements based on each sequence by drawing regions of interest on each of four axial slices through the mid-liver. We compared mDixon-based and GRE-based T2* measurements using paired t-tests and assessed agreement using single-measure intra-class correlation coefficients and Bland-Altman difference plots.
RESULTS: One hundred nine patients met inclusion criteria (site 1=82; site 2=27). Mean age was 12.4±5.8 years, and 42 subjects (39%) were female. There was no statistically significant difference in mean T2* values for the two sequences (pooled means: 11.7±11.0 [GRE] vs. 11.7±10.9 ms [mDixon]; P=0.93). There was excellent absolute agreement between sequences (intraclass correlation coefficient [ICC]=0.98 for patients at both sites, confidence interval [CI]: 0.97-0.98 with mean bias of 0.0 ms [-4.2 ms to +4.2 ms]).
CONCLUSION: 3D mDixon is accurate for measuring liver T2* and can likely replace 2D GRE-based relaxometry.
引用
- Serai SD, Smith EA, Trout AT, et al. Agreement between manual relaxometry and semi-automated scanner-based multi-echo Dixon technique for measuring liver T2* in a pediatric and young adult population. Pediatr Radiol. 2017. PMID: 29058039. doi:10.1007/s00247-017-3990-y