在离体牛肝中使用回声去相关成像实时时空控制高强度聚焦超声热消融

Real-Time Spatiotemporal Control of High-Intensity Focused Ultrasound Thermal Ablation Using Echo Decorrelation Imaging in ex Vivo Bovine Liver.
PubMed 文献 · Ultrasound Med Biol · 2017-10-23

在离体牛肝中评估回声去相关成像反馈控制高强度聚焦超声热消融的能力,当感兴趣区内最小累积回声去相关超过消融控制阈值(-2.7,对应局部消融预测90%特异度)时自动停止声照射。控制组与2、5、9个声周期非控制组比较,控制组病灶面积、宽度和治疗时间显著小于5或9周期非控制组,且预测能力显著优于2周期非控制组。

要点

  • 回声去相关成像可以实时反馈控制HIFU消融,阈值-2.7对应90%特异度。
  • 控制组病灶面积、宽度和治疗时间显著小于5或9周期非控制组。
  • 控制组消融预测能力显著优于2周期非控制组。

核心结论

  • 回声去相关成像似乎是实时HIFU消融控制的有效方法。

研究设计与样本

  • 离体牛肝实验研究,比较控制性HIFU热消融与采用2、5或9个声周期的非控制性HIFU热消融,计算病灶宽度、面积、深度及预测消融的ROC曲线。

诊断标准与关键数值

  • 消融控制阈值由初步实验确定为-2.7(对数尺度去相关每毫秒),对应局部消融预测90%特异度;当感兴趣区内最小累积回声去相关超过该阈值时自动停止声照射。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The ability to control high-intensity focused ultrasound (HIFU) thermal ablation using echo decorrelation imaging feedback was evaluated in ex vivo bovine liver. Sonications were automatically ceased when the minimum cumulative echo decorrelation within the region of interest exceeded an ablation control threshold, determined from preliminary experiments as -2.7 (log-scaled decorrelation per millisecond), corresponding to 90% specificity for local ablation prediction. Controlled HIFU thermal ablation experiments were compared with uncontrolled experiments employing two, five or nine sonication cycles. Means and standard errors of the lesion width, area and depth, as well as receiver operating characteristic curves testing ablation prediction performance, were computed for each group. Controlled trials exhibited significantly smaller average lesion area, width and treatment time than five-cycle or nine-cycle uncontrolled trials and also had significantly greater prediction capability than two-cycle uncontrolled trials. These results suggest echo decorrelation imaging is an effective approach to real-time HIFU ablation control.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Abbass MA, Killin JK, Mahalingam N, et al. Real-Time Spatiotemporal Control of High-Intensity Focused Ultrasound Thermal Ablation Using Echo Decorrelation Imaging in ex Vivo Bovine Liver. Ultrasound Med Biol. 2017. PMID: 29074273. doi:10.1016/j.ultrasmedbio.2017.09.007

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。