评估编码激励在估计热致声速变化中的实施效果

Assessment of Coded Excitation Implementation for Estimating Heat-Induced Speed of Sound Changes.
PubMed 文献 · Ultrasound Med Biol · 2017-10-20

本研究通过计算机模拟、琼脂-明胶体模和离体牛肝实验,比较二进制相位调制Barker序列、线性调频chirp与高斯脉冲在脉冲回波监测热致声速变化中的表现。模拟显示编码激励在极低信噪比下有益,Barker码在0 dB时优于chirp和高斯脉冲;离体肝实验中Barker和chirp均优于高斯脉冲;结论为编码激励在低信噪比下有助于脉冲回波法评估温度诱导的声速变化。

要点

  • 低信噪比(<10 dB)下Barker编码激励性能优于chirp和高斯脉冲。
  • 离体牛肝实验中Barker和chirp的R²均高于高斯脉冲。
  • 高信噪比下编码激励与高斯脉冲性能统计学相当。

核心结论

  • 编码激励可在低信噪比条件下改善脉冲回波法对温度诱导声速变化的评估。

研究设计与样本

  • 研究采用计算机模拟、琼脂-明胶体模和离体牛肝实验,比较二进制相位调制Barker序列、线性频率调制chirp与高斯脉冲传输。
  • 声速变化通过治疗超声系统加热标本实验性诱导,以透射法测得的归一化声速为参考,通过相对回波位移与归一化声速的相关性评估各传输信号性能。

关键影像表现/征象

  • 计算机模拟显示编码激励在极低信噪比下有益,Barker码在信噪比<10 dB时表现优于chirp和高斯脉冲。

诊断标准与关键数值

  • 在0 dB信噪比下,Barker码、chirp和高斯脉冲的R²分别为0.81±0.06、0.68±0.07和0.55±0.08。
  • 离体牛肝实验中,Barker码、chirp和高斯脉冲的R²分别为0.80±0.15、0.79±0.15和0.54±0.17。
  • 琼脂-明胶体模中Barker和chirp与高斯脉冲性能相当。
  • 高信噪比下编码激励与高斯脉冲性能统计学相当。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Speed of sound (SoS) is an acoustic property that is highly sensitive to changes in tissues. SoS can be mapped non-invasively using ultrasonic through transmission wave tomography. This however, practically limits its clinical use to the breast. A pulse-echo-based method that has broader clinical use and that can reliably measure treatment-induced changes in SoS even under poor signal-to-noise ratio (SNR) is highly desirable. The aim of this study was to evaluate the implementation of coded excitations (CoEs) to improve pulse-echo monitoring of heat-induced changes in the SoS. In this study, a binary phase modulated Barker sequence and a linear frequency-modulated chirp were compared with a common Gaussian pulse transmission. The comparison was conducted using computer simulations, as well as transmissions in both agar-gelatin phantoms and ex vivo bovine liver. SoS changes were experimentally induced by heating the specimens with a therapeutic ultrasound system. The performance of each transmission signal was evaluated by correlating the relative echo shifts to the normalized SoS measured by through transmission. The computer simulations indicated that CoEs are beneficial at very low SNR. The Barker code performed better than both the chirp and Gaussian pulses, particularly at SNRs <10 dB (R2 = 0.81 ± 0.06, 0.68 ± 0.07 and 0.55 ± 0.08, respectively, at 0 dB). At high SNRs, the CoEs performed statistically on par with the Gaussian pulse. The experimental findings indicated that both Barker and chirp codes performed better than the Gaussian pulse on ex vivo liver (R2 = 0.80 ± 0.15, 0.79 ± 0.15 and 0.54 ± 0.17, respectively) and comparably on agar-gelatin phantoms. In conclusion, CoEs can be beneficial for assessing temperature-induced changes in the SoS using the pulse-echo method under poor SNR.