文献速递腹部泌尿生殖MR

压缩感知加速三维磁共振胰胆管成像:在疑似胰腺疾病中的应用

Compressed-Sensing Accelerated 3-Dimensional Magnetic Resonance Cholangiopancreatography: Application in Suspected Pancreatic Diseases.
PubMed 文献 · Invest Radiol · 2018-01-01

这项前瞻性研究纳入80例疑似胰腺疾病患者,在3.0 T下随机顺序完成压缩感知屏气、压缩感知导航触发和常规导航触发三种三维磁共振胰胆管成像方案,比较图像质量、胆胰管显示及对管腔相关病变的诊断效能。压缩感知导航触发方案对胰管显示及多种管腔异常的诊断表现最优,Az值0.943~0.983。

要点

  • 压缩感知屏气方案采集时间17秒,压缩感知导航触发方案134.1±33.5秒,均显著短于常规导航触发方案的364.7±78.4秒(均P<0.01)。
  • 压缩感知屏气方案伪影显著少于压缩感知导航触发和常规导航触发方案(均P<0.01)。
  • 压缩感知导航触发方案对胰管近、中、远段显示优于压缩感知屏气方案(均P<0.05),且对管腔异常的诊断效能最高(平均Az值0.943~0.983)。

核心结论

  • 压缩感知磁共振胰胆管成像在疑似胰腺疾病患者中可行,压缩感知导航触发方案对管腔相关病变的诊断准确度更优。

研究设计与样本

  • 前瞻性研究,纳入80例疑似胰腺疾病患者(47男、33女,中位年龄57岁,范围24~87岁),在3.0 T下按随机顺序完成压缩感知屏气、压缩感知导航触发和常规导航触发三种三维磁共振胰胆管成像方案。

关键影像表现/征象

  • 压缩感知屏气方案的伪影显著少于压缩感知导航触发方案和常规导航触发方案(均P<0.01)。
  • 三种方案对胆管显示相当,而压缩感知导航触发方案和常规导航触发方案对胰管近段、中段、远段的显示均优于压缩感知屏气方案(均P<0.05)。

诊断标准与关键数值

  • 图像质量和管腔显示由2名对方案不知情的放射科医师按随机顺序采用5分制独立评分。
  • 采用受试者工作特征曲线及曲线下面积(Az值)比较各方案对管腔相关病变的诊断效能。
  • 压缩感知导航触发方案检测管腔异常、长段管腔狭窄、异常分支胰管及囊性病变与胰管交通的平均Az值为0.943~0.983。

临床应用与阅片要点

  • 在疑似胰腺疾病行三维磁共振胰胆管成像时,压缩感知屏气方案可显著缩短采集时间并减少伪影。
  • 若需精细评估胰管及管腔相关病变,压缩感知导航触发方案对胰管各段显示及诊断效能更优。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aims of this study were to prospectively evaluate image quality, duct visibility, and diagnostic performance in duct-related pathologies of compressed-sensing (CS) accelerated 3-dimensional (3D) magnetic resonance cholangiopancreatography (MRCP) prototype protocols and compare these with those of conventional 3D MRCP protocol in patients with suspected pancreatic diseases.

MATERIAL AND METHODS: The institutional review board approved this prospective study and all patients provided written informed consent. A total of 80 patients (47 men and 33 women; median age, 57 years; age range, 24-87 years) underwent 3D MRCP at 3.0 T. Three protocols were performed in each patient in random order: CS breath-hold (BH) protocol, CS navigator-triggered (NT) protocol, and conventional NT protocol. The acquisition time of each protocol was recorded. Image quality and duct visibility were independently rated in random order on a 5-point scale by 2 radiologists, who were blinded to the protocols. Receiver operating characteristic curves were generated, and area under the curve (Az value) was used to compare the diagnostic performance of each protocol in duct-related pathologies.

RESULTS: Acquisition time was 17 seconds for the CS-BH and 134.1 ± 33.5 seconds for the CS-NT protocol, both being significantly shorter than the conventional NT protocol (364.7 ± 78.4 seconds; both P < 0.01). The CS-BH MRCP protocol showed significantly less artifacts compared with the CS-NT and conventional NT protocols (both P < 0.01). Visualization of bile ducts was comparable in all 3 protocols, whereas CS-NT and conventional NT MRCP depicted pancreatic duct better than CS-BH MRCP did (for proximal, middle, and distal segment; all P < 0.05). Compressed-sensing-NT MRCP had the highest diagnostic performance for detecting ductal anomalies, long-segment duct stenosis, abnormal branch ducts, and communication between cystic lesion and pancreatic duct (mean Az value, 0.943-0.983).

CONCLUSIONS: Compressed-sensing MRCP is feasible in patients with suspected pancreatic diseases. Compressed-sensing-NT MRCP demonstrated superior diagnostic accuracy for duct-related pathologies.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Zhu L, Wu X, Sun Z, et al. Compressed-Sensing Accelerated 3-Dimensional Magnetic Resonance Cholangiopancreatography: Application in Suspected Pancreatic Diseases. Invest Radiol. 2018. PMID: 28976478. doi:10.1097/RLI.0000000000000421

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。