腹部CT对比增强:不同扫描仪光子能量的影响

Contrast enhancement in abdominal computed tomography: influence of photon energy of different scanners.
PubMed 文献 · Br J Radiol · 2017-11-06

研究通过体模和临床比较两台不同厂商CT扫描仪的对比增强差异。体模中扫描仪B在80、100、120 kVp下对比增强分别比扫描仪A高36.9%、32.6%、30.8%,扫描仪A有效能量更高。临床定量与定性分析均显示扫描仪B对肝实质、胰腺、肾、门静脉及下腔静脉的增强显著优于扫描仪A。

要点

  • 同扫描及对比参数下,不同厂商CT扫描仪腹部脏器对比增强存在显著差异。
  • 扫描仪B的对比增强在体模各kVp及临床腹部脏器均显著高于扫描仪A。
  • 解读临床CT图像时需注意不同X射线能量对对比剂衰减的影响。

核心结论

  • 两台不同厂商CT扫描仪在相同扫描和对比参数下,腹部脏器对比增强存在显著差异,扫描仪B增强效果优于扫描仪A。

研究设计与样本

  • 体模研究比较两台扫描仪在80、100、120 kVp的对比增强并计算有效能量;临床研究纳入扫描仪A组40例和扫描仪B组各40例患者,使用相同扫描方案行腹部扫描。

关键影像表现/征象

  • 体模研究中扫描仪B的对比增强在80、100、120 kVp下分别比扫描仪A高36.9%、32.6%、30.8%。
  • 扫描仪A的有效能量高于扫描仪B。
  • 临床定量分析显示扫描仪B对肝实质、胰腺、肾、门静脉和下腔静脉的对比增强显著优于扫描仪A。
  • 扫描仪B图像的对比增强平均视觉评分也显著高于扫描仪A。

临床应用与阅片要点

  • 不同厂商CT扫描仪即使采用相同扫描和对比参数,腹部脏器对比增强也可能不同。
  • 解读临床CT图像时需注意不同X射线能量对对比剂衰减的影响。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Different CT scanners have different X-ray spectra and photon energies indicating that contrast enhancement vary among scanners. However, this issue has not been fully validated; therefore, we performed phantom and clinical studies to assess this difference.

METHODS: Two scanners were used: scanner-A and scanner-B. In the phantom study, we compared the contrast enhancement between the scanners at tube voltage peaks of 80, 100 and 120 kVp. Then, we calculated the effective energies of the two CT scanners. In the clinical study, 40 patients underwent abdominal scanning with scanner-A and another 40 patients with scanner-B, with each group using the same scanning protocol. The contrast enhancement of abdominal organs was assessed quantitatively (based on the absolute difference between the attenuation of unenhanced scans and contrast-enhanced scans) and qualitatively. A two-tailed independent Student's t-test and or the Mann-Whitney U test were used to compare the discrepancies.

RESULTS: In the phantom study, contrast enhancement for scanner-B was 36.9, 32.6 and 30.8% higher than that for scanner-A at 80, 100 and 120 kVp, respectively. The effective energies were higher for scanner-A than for scanner-B. In the quantitative analysis for the clinical study, scanner-B yielded significantly better contrast enhancement of the hepatic parenchyma, pancreas, kidney, portal vein and inferior vena cava compared with that of scanner-A. The mean visual scores for contrast enhancement were also significantly higher on images obtained by scanner-B than those by scanner-A.

CONCLUSION: There were significant differences in contrast enhancement of the abdominal organs between the compared CT scanners from two different vendors even at the same scanning and contrast parameters. Advances in knowledge: Awareness of the impact of different X-ray energies on the resultant attenuation of contrast material is important when interpreting clinical CT images.