二维剪切波弹性成像(GE系统)与虚拟触诊组织量化(Siemens系统)在肝纤维化分级中的个体内比较
该研究对93例各种病因肝病患者行个体内比较,评估GE二维剪切波弹性成像与Siemens虚拟触诊组织量化能否互换用于肝纤维化分级。两种技术各自重复性优良(组内相关系数GE为0.89、VTTQ为0.90),但两法间分级一致性仅为一般(κ=0.31),提示不宜互换使用。
要点
- GE二维剪切波弹性成像与Siemens虚拟触诊组织量化各自重复性优良,组内相关系数分别为0.89与0.90。
- 两法间肝纤维化分级一致性仅为一般(κ=0.31),提示不宜互换使用。
- 应用已发表的系统特异性剪切波速度阈值时,两法对F0/F1与≥F2的区分一致性有限。
核心结论
- 两种技术各自重复性良好至优秀,但两法间肝纤维化分级一致性仅为一般(κ=0.31),提示不宜互换使用。
研究设计与样本
- 个体内比较研究,纳入93例各种病因肝病患者(男51例、女42例,平均年龄54岁),采用GE系统二维剪切波弹性成像与Siemens系统虚拟触诊组织量化测量,以已发表的系统特异性剪切波速度阈值将肝纤维化分为临床非显著(F0/F1)与显著(≥F2)。
诊断标准与关键数值
- 采用已发表的系统特异性剪切波速度阈值将肝纤维化分为临床非显著(F0/F1)与显著(≥F2);组内重复性组内相关系数GE二维剪切波弹性成像为0.89(0.84-0.93)、VTTQ为0.90(0.86-0.93);组内分级一致性κ值GE为0.65、VTTQ为0.82;两法间分级一致性κ=0.31。
临床应用与阅片要点
- 作者认为,尽管两种技术各自重复性良好至优秀,但两法间分级一致性仅为一般,提示不宜互换使用。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Ultrasound-based shear wave elastography (SWE) has recently gained substantial attention for non-invasive assessment of liver fibrosis. The purpose of this study was to perform an intra-individual comparison between 2-D shear wave elastography (2-D-SWE with a GE system) and Virtual Touch Tissue Quantification (VTTQ with a Siemens system) to assess whether these can be used interchangeably to grade fibrosis. Ninety-three patients (51 men, 42 women; mean age, 54 y) with liver disease of various etiologies (hepatitis B virus = 47, hepatitis C virus = 22; alcohol = 6, non-alcoholic steatohepatitis = 5, other = 13) were included. Using published system-specific shear wave speed cutoff values, liver fibrosis was classified into clinically non-significant (F0/F1) and significant (≥F2) fibrosis. Results indicated that intra-modality repeatability was excellent for both techniques (GE 2-D-SWE: intra-class correlation coefficient = 0.89 [0.84-0.93]; VTTQ: intra-class correlation coefficient = 0.90 [0.86-0.93]). Intra-modality classification agreement for fibrosis grading was good to excellent (GE 2-D-SWE: κ = 0.65, VTTQ: κ = 0.82). However, inter-modality agreement for fibrosis grading was only fair (κ = 0.31) using published system-specific shear wave speed cutoff values of fibrosis. In conclusion, although both GE 2-D-SWE and Siemens VTTQ exhibit good to excellent intra-modality repeatability, inter-modality agreement is only fair, suggesting that these should not be used interchangeably.
引用
- Sigrist RMS, El Kaffas A, Jeffrey RB, et al. Intra-Individual Comparison between 2-D Shear Wave Elastography (GE System) and Virtual Touch Tissue Quantification (Siemens System) in Grading Liver Fibrosis. Ultrasound Med Biol. 2017. PMID: 28967501. doi:10.1016/j.ultrasmedbio.2017.08.934
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。