文献速递心血管技术与安全MR

3特斯拉磁共振成像后人类外周血单个核细胞中DNA双链断裂缺失:γH2AX流式细胞术评估

Absence of DNA double-strand breaks in human peripheral blood mononuclear cells after 3 Tesla magnetic resonance imaging assessed by γH2AX flow cytometry.
PubMed 文献 · Eur Radiol · 2017-10-06

本研究采用γH2AX染色和流式细胞术,评估3-T心脏磁共振检查对外周血单个核细胞DNA双链断裂的影响。73名健康非吸烟志愿者中,36名接受心脏MRI,37名对照无干预。两组γH2AX信号从基线至检查后均显著增加(分别6.7%和7.8%),但组间差异无统计学意义,结论为3-T MRI未损害DNA完整性。

要点

  • 3-T心脏MRI后外周血单个核细胞γH2AX信号未显示DNA双链断裂证据
  • 两组γH2AX信号较基线均显著升高,但组间差异无统计学意义
  • 前瞻性对照试验纳入男女两性,支持MRI在DNA损伤方面安全使用

核心结论

  • 基于本研究,γH2AX流式细胞术未显示3-T心脏MRI检查损害外周血单个核细胞DNA完整性。

研究设计与样本

  • 前瞻性对照试验,73名健康非吸烟志愿者,36名接受心脏MRI(CMRI),37名对照无干预,采用γH2AX染色和流式细胞术分析,混合线性模型重复测量分析组间差异。

诊断标准与关键数值

  • γH2AX信号从基线至检查后:CMRI组增加6.7%(SD 7.18%),对照组增加7.8%(SD 6.61%),组间差异无统计学意义。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Magnetic resonance imaging (MRI) is regarded as a non-harming and non-invasive imaging modality with high tissue contrast and almost no side effects. Compared to other cross-sectional imaging modalities, MRI does not use ionising radiation. Recently, however, strong magnetic fields as applied in clinical MRI scanners have been suspected to induce DNA double-strand breaks in human lymphocytes.

METHODS: In this study we investigated the impact of 3-T cardiac MRI examinations on the induction of DNA double-strand breaks in peripheral mononuclear cells by γH2AX staining and flow cytometry analysis. The study cohort consisted of 73 healthy non-smoking volunteers with 36 volunteers undergoing CMRI and 37 controls without intervention. Differences between the two cohorts were analysed by a mixed linear model with repeated measures.

RESULTS: Both cohorts showed a significant increase in the γH2AX signal from baseline to post-procedure of 6.7 % (SD 7.18 %) and 7.8 % (SD 6.61 %), respectively. However, the difference between the two groups was not significant.

CONCLUSION: Based on our study, γH2AX flow cytometry shows no evidence that 3-T MRI examinations as used in cardiac scans impair DNA integrity in peripheral mononuclear cells.

KEY POINTS: • No evidence for DNA double-strand breaks after cardiac MRI. • Prospective study underlines safe use of MRI with regard to DNA damage. • Controlled trial involving both genders investigating DNA DSBs after 3-T MRI.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Fasshauer M, Krüwel T, Zapf A, et al. Absence of DNA double-strand breaks in human peripheral blood mononuclear cells after 3 Tesla magnetic resonance imaging assessed by γH2AX flow cytometry. Eur Radiol. 2017. PMID: 28986695. doi:10.1007/s00330-017-5056-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。