新一代自适应统计迭代重建(ASIR-V)在胸部CT中降低噪声与提升图像质量的评估

Assessment of noise reduction potential and image quality improvement of a new generation adaptive statistical iterative reconstruction (ASIR-V) in chest CT.
PubMed 文献 · Br J Radiol · 2017-11-16

研究纳入30例接受常规剂量平扫胸部CT的患者,比较40%ASIR与10%–100%ASIR-V重建的图像质量。ASIR-V比例升高时客观噪声下降、SNR升高;30%–100%ASIR-V的SD显著低于40%ASIR;主观上50%–70%ASIR-V显示小结构优于40%ASIR及其他比例,60%ASIR-V图像质量最高,80%及以上ASIR-V图像锐利度显著下降。

要点

  • 30%–100%ASIR-V的客观噪声(SD)显著低于40%ASIR(p < 0.05)。
  • 50%–70%ASIR-V显示小结构显著优于40%ASIR及其他比例ASIR-V(p < 0.05)。
  • 60%ASIR-V在全部ASIR-V重建中主观图像质量最高。

核心结论

  • 在常规剂量平扫胸部CT中,与推荐的40%ASIR相比,ASIR-V在降低图像噪声和伪影、保持图像锐利度方面更具潜力;综合客观与主观评价,60%ASIR-V重建的图像质量最高。

研究设计与样本

  • 临床研究,纳入30例接受常规剂量平扫胸部CT的患者,使用GE Discovery CT750HU扫描仪,管电压120 kVp、自动管电流调制噪声指数14HU、转速0.6 s、螺距1.375:1、层厚5 mm;所有扫描以推荐的40%ASIR及10%–100%ASIR-V(10%递增)重建。

诊断标准与关键数值

  • 在气管隆突层面测量皮下脂肪、背部肌肉和降主动脉的CT衰减值和SD,以SD代表图像噪声并计算SNR;两名有经验的放射科医师独立进行主观图像质量评价。

临床应用与阅片要点

  • 作者结论提示ASIR-V较ASIR算法提供更好的图像质量和更高的诊断信心;综合主客观评价,平扫胸部CT以60%ASIR-V重建的图像质量最高。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the image quality improvement and noise reduction in routine dose, non-enhanced chest CT imaging by using a new generation adaptive statistical iterative reconstruction (ASIR-V) in comparison with ASIR algorithm.

METHODS: 30 patients who underwent routine dose, non-enhanced chest CT using GE Discovery CT750HU (GE Healthcare, Waukesha, WI) were included. The scan parameters included tube voltage of 120 kVp, automatic tube current modulation to obtain a noise index of 14HU, rotation speed of 0.6 s, pitch of 1.375:1 and slice thickness of 5 mm. After scanning, all scans were reconstructed with the recommended level of 40%ASIR for comparison purpose and different percentages of ASIR-V from 10% to 100% in a 10% increment. The CT attenuation values and SD of the subcutaneous fat, back muscle and descending aorta were measured at the level of tracheal carina of all reconstructed images. The signal-to-noise ratio (SNR) was calculated with SD representing image noise. The subjective image quality was independently evaluated by two experienced radiologists.

RESULTS: For all ASIR-V images, the objective image noise (SD) of fat, muscle and aorta decreased and SNR increased along with increasing ASIR-V percentage. The SD of 30% ASIR-V to 100% ASIR-V was significantly lower than that of 40% ASIR (p < 0.05). In terms of subjective image evaluation, all ASIR-V reconstructions had good diagnostic acceptability. However, the 50% ASIR-V to 70% ASIR-V series showed significantly superior visibility of small structures when compared with the 40% ASIR and ASIR-V of other percentages (p < 0.05), and 60% ASIR-V was the best series of all ASIR-V images, with a highest subjective image quality. The image sharpness was significantly decreased in images reconstructed by 80% ASIR-V and higher.

CONCLUSION: In routine dose, non-enhanced chest CT, ASIR-V shows greater potential in reducing image noise and artefacts and maintaining image sharpness when compared to the recommended level of 40%ASIR algorithm. Combining both the objective and subjective evaluation of images, non-enhanced chest CT images reconstructed with 60% ASIR-V have the highest image quality. Advances in knowledge: This is the first clinical study to evaluate the clinical value of ASIR-V in the same patients using the same CT scanner in the non-enhanced chest CT scans. It suggests that ASIR-V provides the better image quality and higher diagnostic confidence in comparison with ASIR algorithm.

来源:Br J Radiol,查看原文。版权归原作者所有。